Skip to the content

FDA Medical Device Recalls (openFDA)

Limacorporate S.p.A: PRIMA TT Glenoid Monoblock/Modular reverse TT baseplates: Description/Product Code: BASEPLATE D.25MM REGULAR/1975.14.500, BASEPLATE D.28MM REGULAR/1975.14.800, BASEPLATE D.25MM FULL WEDGED 10-degrees/1975.14.510, BASEPLATE D.28MM FULL WEDGED 10-degrees/1975.14.810, BASEPLATE D.25MM FULL WEDGED 15-degrees/1975.14.615, BASEPLATE D.28MM FULL WEDGED 15-degrees/1975.14.815, BASEPLATE D.28MM FULL WEDGED 15-degrees X/1975.14.865, BASEPLATE D.25 MM FULL WEDGED 20-degrees/1975.14.620, BASEPLATE D.28 MM FULL WEDGED 20-degrees/1975.14.820, BASEPLATE D.28 MM FULL WEDGED 20-degrees X/1975.14.870

Recall number
Z-0726-2024
Recalling firm
Limacorporate S.p.A
Product
PRIMA TT Glenoid Monoblock/Modular reverse TT baseplates: Description/Product Code: BASEPLATE D.25MM REGULAR/1975.14.500, BASEPLATE D.28MM REGULAR/1975.14.800, BASEPLATE D.25MM FULL WEDGED 10-degrees/1975.14.510, BASEPLATE D.28MM FULL WEDGED 10-degrees/1975.14.810, BASEPLATE D.25MM FULL WEDGED 15-degrees/1975.14.615, BASEPLATE D.28MM FULL WEDGED 15-degrees/1975.14.815, BASEPLATE D.28MM FULL WEDGED 15-degrees X/1975.14.865, BASEPLATE D.25 MM FULL WEDGED 20-degrees/1975.14.620, BASEPLATE D.28 MM FULL WEDGED 20-degrees/1975.14.820, BASEPLATE D.28 MM FULL WEDGED 20-degrees X/1975.14.870
Status
Open, Classified
Initiated
2023-08-29
Posted
2024-01-12
Terminated
not on file
Reason
Potential for baseplate peripheral holes to be out of specification due to a manufacturing issue, which could lead to loosening of the implant, or the screw moving from its original position.
Root cause
Device Design
Action
On 8/29/23, the firm phoned all agents and surgeons to immediately stop surgeries with affected devices. On 10/10/23 the firm emailed recall notices to customers asking them to take the following actions: 1) Surgeons are suggested to perform a short-term patient s monitoring (radiographs at 6 weeks, 3-6 months and 1 year might be considered) and acting accordingly, based on a case-by-case evaluation. 2) Locate, quarantine, and return affected devices. Complete and return the response form to uscomplaint@limacorporate.com 3) Customers with questions can email the firm at medicalcomplaints@limacorporate.com
Quantity
179
Distribution
US: IL, MO, CA, IN, MI, OK
Lot, serial or model codes
Product Code/UDI-DI: 1975.14.500/08033390244824, 1975.14.800/08033390244893, 1975.14.510/08033390231299, 1975.14.810/08033390244909, 1975.14.615/08033390244848, 1975.14.815/08033390244916, 1975.14.865/08033390244930, 1975.14.620/08033390231312, 1975.14.820/08033390244923, 1975.14.870/08033390244947
510(k) numbers
["K222427"]
PMA numbers
not on file
Product code
MBF
Firm FEI
3008021110
Firm city
San Daniele Del Friuli
Firm state
not on file
Firm country
Italy

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register