Skip to the content

FDA Medical Device Recalls (openFDA)

Smiths Medical ASD Inc.: PORTEX Bivona Cuffless Neonatal/Pediatric FlexTend Tracheostomy Tube, REF 60PFSS45

Recall number
Z-0726-2021
Recalling firm
Smiths Medical ASD Inc.
Product
PORTEX Bivona Cuffless Neonatal/Pediatric FlexTend Tracheostomy Tube, REF 60PFSS45
Status
Terminated
Initiated
2020-12-15
Posted
not on file
Terminated
2022-12-20
Reason
A specific model and lot number of Bivona Cuffless FlexTend Tracheostomy Tube may have been incorrectly packaged with a V-neck flange instead of a straight flange.
Root cause
Packaging process control
Action
The firm initiated the recall by letter on 12/15/2020. The letter requested the following: "INSTRUCTIONS TO CUSTOMERS AND DISTRIBUTORS: 1. Locate affected product in your possession by referring to the Affected Devices List (Table 1) listed on this Recall Notice. 2. Determine the number of affected devices in your possession and complete the Response Form (Attachment 1) within 10 days of receipt, returning it to fieldactions@smiths-medical.com. The form must be returned even if you do not have any affected Bivona Cuffless FlexTend Tracheostomy Tube Devices in your possession. 3. All affected product must be returned to Smiths Medical for processing. A pre-paid return shipping label will be sent to you when the Response Form is returned indicating whether impacted devices are in your possession. When returning product please include a copy of the Response Form (Attachment 1) inside EACH BOX. 4. Product credit will be processed once the impacted product and Response Form (Attachment 1) has been received and processed. 5. DISTRIBUTORS: if you have distributed potentially affected product to your customers, please immediately notify them of this Recall with the editable Response Form for your customers (Attachment 1a) to respond appropriately."
Quantity
238 devices
Distribution
US nationwide distribution.
Lot, serial or model codes
Lot Number 3991195
510(k) numbers
["K944178"]
PMA numbers
not on file
Product code
JOH
Firm FEI
3012307300
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register