Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott Laboratories HPD/ADD: Abbott IMx AFP Calibrators, list 2271-01; a box of 6 bottles (4 mL each) with AFP (human), nonreactive for anti-HIV-1/HIV-2 and HCV and nonreactive for HBsAg, inbuffered calf serum at the following concentrations: A - 0, B - 15, C - 50, D - 100, E - 200 and F - 350; Abbott Diagnostics, a Division of Abbott Laboratories, Abbott park, IL 60064

Recall number
Z-0726-03
Recalling firm
Abbott Laboratories HPD/ADD
Product
Abbott IMx AFP Calibrators, list 2271-01; a box of 6 bottles (4 mL each) with AFP (human), nonreactive for anti-HIV-1/HIV-2 and HCV and nonreactive for HBsAg, inbuffered calf serum at the following concentrations: A - 0, B - 15, C - 50, D - 100, E - 200 and F - 350; Abbott Diagnostics, a Division of Abbott Laboratories, Abbott park, IL 60064
Status
Terminated
Initiated
2003-03-04
Posted
2003-04-16
Terminated
2003-09-04
Reason
The IMx AFP Calibrators may cause depressed values for the IMx AFP Low Control, and cause patient results to be 20% lower than actual values
Root cause
Other
Action
Recalled by letter dated 3/4/03. The laboratory accounts were informed of the lot of calibrators causing depressed values for the IMx AFP Low Control, requested to destroy any of the suspect lot of calibrators, and recalibrate the AFP assay using a different lot of calibrators.
Quantity
107 kits
Distribution
Nationwide
Lot, serial or model codes
lot number 89651M300, exp. 04/20/03
510(k) numbers
not on file
PMA numbers
["P820060"]
Product code
LOJ
Firm FEI
1415939
Firm city
Abbott Park
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register