Enterix, Inc.: InSure ONE Fecal Immunochemical Test. Labeled as follows: 1. Insure ONE POL Patient Kit - 10 pack. Model Number: 90010. 2. Insure ONE POL Patient Kit - 25 Pack. Model Number: 90025. 3. InSure ONE POL Combo Pack. Model Number: 90030. InSure ONE is a fecal immunochemical test (FIT) that qualitatively detects human hemoglobin from blood in fecal samples.

Recall number
Z-0725-2026
Recalling firm
Enterix, Inc.
Product
InSure ONE Fecal Immunochemical Test. Labeled as follows: 1. Insure ONE POL Patient Kit - 10 pack. Model Number: 90010. 2. Insure ONE POL Patient Kit - 25 Pack. Model Number: 90025. 3. InSure ONE POL Combo Pack. Model Number: 90030. InSure ONE is a fecal immunochemical test (FIT) that qualitatively detects human hemoglobin from blood in fecal samples.
Status
Open, Classified
Initiated
2025-09-17
Posted
2025-11-28
Terminated
not on file
Reason
InSure ONE failed sensitivity testing when used with Test Card Lot T241102, which may result in false negative results.
Root cause
Under Investigation by firm
Action
Consignees were initially sent an email as notification of the recall on about 09/17/2025, which was followed by a formal recall notification letter on about 09/23/2025. Consignees were instructed to immediately examine inventory and quarantine any affected units, notify customers if affected further distributed, and for those patients that have already been tested using an InSure ONE test kit that has been identified as part of the scope of this recall be retested. Once an inventory count has been received by Enterix, they will provide RMA and courier information so the recalled product can be returned for sorting and destruction.
Quantity
201,426 test kits
Distribution
Worldwide - US Nationwide distribution in the states of California, Florida, Georgia, Illinois, Indiana, Kentucky, Massachusetts, Nevada, New Jersey, New York, North Carolina, Ohio, Pennsylvania, Texas, Washington, West Virginia and the country of Philippines.
Lot, serial or model codes
1. Model Number: 90010; GTIN:10850093003313; All lots produced from 7/23/2025-9/12/2025 with Test card lot TT241102. 2. Model Number: 90025; GTIN: 20850093003310; All lots produced from 7/23/2025-9/12/2025 with Test card lot TT241102. 3. Model Number: 90030; GTIN: 00850093003279; All lots produced from 7/23/2025-9/12/2025 with Test card lot TT241102.
510(k) numbers
["K170548"]
PMA numbers
not on file
Product code
KHE
Firm FEI
3003224146
Firm city
Edison
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register