FDA Medical Device Recalls (openFDA)
Enterix, Inc.: InSure ONE Fecal Immunochemical Test. Labeled as follows: 1. Insure ONE POL Patient Kit - 10 pack. Model Number: 90010. 2. Insure ONE POL Patient Kit - 25 Pack. Model Number: 90025. 3. InSure ONE POL Combo Pack. Model Number: 90030. InSure ONE is a fecal immunochemical test (FIT) that qualitatively detects human hemoglobin from blood in fecal samples.
- Recall number
- Z-0725-2026
- Recalling firm
- Enterix, Inc.
- Product
- InSure ONE Fecal Immunochemical Test. Labeled as follows: 1. Insure ONE POL Patient Kit - 10 pack. Model Number: 90010. 2. Insure ONE POL Patient Kit - 25 Pack. Model Number: 90025. 3. InSure ONE POL Combo Pack. Model Number: 90030. InSure ONE is a fecal immunochemical test (FIT) that qualitatively detects human hemoglobin from blood in fecal samples.
- Status
- Open, Classified
- Initiated
- 2025-09-17
- Posted
- 2025-11-28
- Terminated
- not on file
- Reason
- InSure ONE failed sensitivity testing when used with Test Card Lot T241102, which may result in false negative results.
- Root cause
- Under Investigation by firm
- Action
- Consignees were initially sent an email as notification of the recall on about 09/17/2025, which was followed by a formal recall notification letter on about 09/23/2025. Consignees were instructed to immediately examine inventory and quarantine any affected units, notify customers if affected further distributed, and for those patients that have already been tested using an InSure ONE test kit that has been identified as part of the scope of this recall be retested. Once an inventory count has been received by Enterix, they will provide RMA and courier information so the recalled product can be returned for sorting and destruction.
- Quantity
- 201,426 test kits
- Distribution
- Worldwide - US Nationwide distribution in the states of California, Florida, Georgia, Illinois, Indiana, Kentucky, Massachusetts, Nevada, New Jersey, New York, North Carolina, Ohio, Pennsylvania, Texas, Washington, West Virginia and the country of Philippines.
- Lot, serial or model codes
- 1. Model Number: 90010; GTIN:10850093003313; All lots produced from 7/23/2025-9/12/2025 with Test card lot TT241102. 2. Model Number: 90025; GTIN: 20850093003310; All lots produced from 7/23/2025-9/12/2025 with Test card lot TT241102. 3. Model Number: 90030; GTIN: 00850093003279; All lots produced from 7/23/2025-9/12/2025 with Test card lot TT241102.
- 510(k) numbers
- ["K170548"]
- PMA numbers
- not on file
- Product code
- KHE
- Firm FEI
- 3003224146
- Firm city
- Edison
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28