FDA Medical Device Recalls (openFDA)
Terumo Medical Corporation: Destination Guiding Sheath catheter introducer, Catalog Number: RSR06 Product Usage: The Destination Guiding Sheath is designed to be used for the introduction of interventional and diagnostic devices into the human vasculature, including but not limited to the renal, carotid, and peripheral arteries.
- Recall number
- Z-0725-2018
- Recalling firm
- Terumo Medical Corporation
- Product
- Destination Guiding Sheath catheter introducer, Catalog Number: RSR06 Product Usage: The Destination Guiding Sheath is designed to be used for the introduction of interventional and diagnostic devices into the human vasculature, including but not limited to the renal, carotid, and peripheral arteries.
- Status
- Terminated
- Initiated
- 2017-09-12
- Posted
- not on file
- Terminated
- 2018-08-31
- Reason
- The tucking depth feature of this catheter introducer is marginally outside the upper specification.
- Root cause
- Process control
- Action
- Terumo sent an Urgent Medical Device Recall dated September 12, 2017 to affected customers. The letter identified the affected product problem and actions to be taken. The informed customers that a territory managers will visit each institution to examine the inventory, identify the affected product and lot, and prepare the product for return to Terumo Medical Corporation. There is an inventory confirmation form that is returned by the sales representative to the recalling site.
- Quantity
- 33
- Distribution
- To be added.
- Lot, serial or model codes
- Lot: VK09, Expiration Date: 31-Jan-2020
- 510(k) numbers
- ["K012584"]
- PMA numbers
- not on file
- Product code
- DYB
- Firm FEI
- 1118880
- Firm city
- Elkton
- Firm state
- MD
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28