Skip to the content

FDA Medical Device Recalls (openFDA)

Terumo Medical Corporation: Destination Guiding Sheath catheter introducer, Catalog Number: RSR06 Product Usage: The Destination Guiding Sheath is designed to be used for the introduction of interventional and diagnostic devices into the human vasculature, including but not limited to the renal, carotid, and peripheral arteries.

Recall number
Z-0725-2018
Recalling firm
Terumo Medical Corporation
Product
Destination Guiding Sheath catheter introducer, Catalog Number: RSR06 Product Usage: The Destination Guiding Sheath is designed to be used for the introduction of interventional and diagnostic devices into the human vasculature, including but not limited to the renal, carotid, and peripheral arteries.
Status
Terminated
Initiated
2017-09-12
Posted
not on file
Terminated
2018-08-31
Reason
The tucking depth feature of this catheter introducer is marginally outside the upper specification.
Root cause
Process control
Action
Terumo sent an Urgent Medical Device Recall dated September 12, 2017 to affected customers. The letter identified the affected product problem and actions to be taken. The informed customers that a territory managers will visit each institution to examine the inventory, identify the affected product and lot, and prepare the product for return to Terumo Medical Corporation. There is an inventory confirmation form that is returned by the sales representative to the recalling site.
Quantity
33
Distribution
To be added.
Lot, serial or model codes
Lot: VK09, Expiration Date: 31-Jan-2020
510(k) numbers
["K012584"]
PMA numbers
not on file
Product code
DYB
Firm FEI
1118880
Firm city
Elkton
Firm state
MD
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register