FDA Medical Device Recalls (openFDA)
Cepheid: Xpert MTB/RIF. Model Number: GXMTB/RIF-US-10.
- Recall number
- Z-0724-2026
- Recalling firm
- Cepheid
- Product
- Xpert MTB/RIF. Model Number: GXMTB/RIF-US-10.
- Status
- Open, Classified
- Initiated
- 2025-11-05
- Posted
- 2025-11-26
- Terminated
- not on file
- Reason
- Product testing did not meet expected stability criteria.
- Root cause
- Under Investigation by firm
- Action
- Cephid notified consignees on about 11/05/2025 via FedEx letter. Consignees were instructed to ensure no affected units remain in inventory or are available for use since the affected lots have passed their expiration date. Past results obtained from Xpert MTB/RIF batch 1000812229 lot 41302 may be inaccurate. If these results were used to make clinical management decisions, it is recommended that the results be reviewed in the context of corresponding culture data, drug susceptibility testing, clinical presentation and other relevant laboratory information to ensure accurate patient care. If there is uncertainty or insufficient information to confirm the result, repeat testing to ensure accuracy. Consignees were requested to notify relevant clinical teams as needed. Cepheid will provide products for the purposes of retesting. Kits are expected to ship within 5 business days following receipt of the completed response form. Additionally, consignees were instructed to complete and return the Customer Response Form.
- Quantity
- 9,880 units
- Distribution
- US Nationwide distribution and Puerto Rico.
- Lot, serial or model codes
- Model Number: GXMTB/RIF-US-10. UDI-DI: 07332940000912. Batch: 1000812229. Lot: 41302. Expiration Date: 2025-02-02
- 510(k) numbers
- ["K143302"]
- PMA numbers
- not on file
- Product code
- PEU
- Firm FEI
- 3004530258
- Firm city
- Sunnyvale
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28