Cepheid: Xpert MTB/RIF. Model Number: GXMTB/RIF-US-10.

Recall number
Z-0724-2026
Recalling firm
Cepheid
Product
Xpert MTB/RIF. Model Number: GXMTB/RIF-US-10.
Status
Open, Classified
Initiated
2025-11-05
Posted
2025-11-26
Terminated
not on file
Reason
Product testing did not meet expected stability criteria.
Root cause
Under Investigation by firm
Action
Cephid notified consignees on about 11/05/2025 via FedEx letter. Consignees were instructed to ensure no affected units remain in inventory or are available for use since the affected lots have passed their expiration date. Past results obtained from Xpert MTB/RIF batch 1000812229 lot 41302 may be inaccurate. If these results were used to make clinical management decisions, it is recommended that the results be reviewed in the context of corresponding culture data, drug susceptibility testing, clinical presentation and other relevant laboratory information to ensure accurate patient care. If there is uncertainty or insufficient information to confirm the result, repeat testing to ensure accuracy. Consignees were requested to notify relevant clinical teams as needed. Cepheid will provide products for the purposes of retesting. Kits are expected to ship within 5 business days following receipt of the completed response form. Additionally, consignees were instructed to complete and return the Customer Response Form.
Quantity
9,880 units
Distribution
US Nationwide distribution and Puerto Rico.
Lot, serial or model codes
Model Number: GXMTB/RIF-US-10. UDI-DI: 07332940000912. Batch: 1000812229. Lot: 41302. Expiration Date: 2025-02-02
510(k) numbers
["K143302"]
PMA numbers
not on file
Product code
PEU
Firm FEI
3004530258
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register