Biomet, Inc.: Vanguard 360 Revision System Posterior Augment With Bolt - Product Usage: intended for use in knee arthroplasty Item Number: 185422

Recall number
Z-0723-2021
Recalling firm
Biomet, Inc.
Product
Vanguard 360 Revision System Posterior Augment With Bolt - Product Usage: intended for use in knee arthroplasty Item Number: 185422
Status
Terminated
Initiated
2020-12-01
Posted
not on file
Terminated
2022-07-08
Reason
Incorrect bolt -tibial bolts in package instead of femoral bolts may delay procedure
Root cause
Process control
Action
.Zimmer Biomet issued Urgent Medical Device Recall letter issued on 12/1/20 to distributors with direct sales and/or fielded inventory via email and traceable letter sent via FedEx. Hospitals with direct sales or current inventory will be notified via FedEx. Letter states reason for recall , health risk and action to take: Distributors esponsibilities include: locating and removing the product in their territory, as well as identifying hospitals who have previously used the product. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the field safety notice for hospitals and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com.
Quantity
21 units
Distribution
Worldwide distribution - US Nationwide distribution in the state of FL and the countries of Korea, Japan, Netherlands.
Lot, serial or model codes
Lot Number: 098780  UDI Number: (01) 00880304482739 (17) 290501 (10) 098780
510(k) numbers
["K093293"]
PMA numbers
not on file
Product code
JWH
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register