MEDLINE INDUSTRIES, LP - Northfield: Medline brand, medical procedure convenience kits, labeled as: 1) MVFF/FIBU #54-RF , SKU DYNJ21927R; 2) DIEP FLAP, SKU DYNJ910120A

Recall number
Z-0721-2025
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
Medline brand, medical procedure convenience kits, labeled as: 1) MVFF/FIBU #54-RF , SKU DYNJ21927R; 2) DIEP FLAP, SKU DYNJ910120A
Status
Open, Classified
Initiated
2024-11-06
Posted
2024-12-20
Terminated
not on file
Reason
Surgical Patties & Surgical Strips in specific lots of Custom Convenience Kits were found to contain higher-than-expected levels of endotoxin within the raw material used to produce sterile Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in those finished goods
Root cause
Nonconforming Material/Component
Action
An IMMEDIATE ACTION REQUIRED MEDICAL DEVICE RECALL notification letter dated 11/7/24 was sent to customers. REQUIRED ACTION: 1.Immediately check your stock for the affected item number and the affected lot numbers which can be located within the recall portal. Quarantine all affected product immediately. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. 3.Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. The login for completing the response form is: Website link: https://recalls.medline.comRecall Reference #: R-24-209-FGX1 Recall Code: Recall Code 4.If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Quantity
318 kits
Distribution
US Nationwide. Canada.
Lot, serial or model codes
UDI-DI: DYNJ21927R, 10195327104733(each), 40195327104734(case); Lot Number: 23IBF547 23GBT501 23FBU360 23EBU929 23EBL624 23EBA114 23CBN730 23ABF393 22JBK203 22IBN858 22HBO301 22GBC794 22CBI774; UDI-DI: DYNJ910120A, 10195327689872 (each), 40195327689873 (case), Lot Numbers: 24HMK254
510(k) numbers
not on file
PMA numbers
not on file
Product code
FDE
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register