Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics, Inc.: ADVIA Gamma-Glutamyl Transferase (GGT) Reagent - Product Usage: For in vitro diagnostic use in the quantitative determination of gamma-glutamyl transferase in human serum and plasma. Siemens Material Number (SMN): ADVIA Chemistry: 10309495 (20 mL) 10316298 (70 mL)

Recall number
Z-0721-2021
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
ADVIA Gamma-Glutamyl Transferase (GGT) Reagent - Product Usage: For in vitro diagnostic use in the quantitative determination of gamma-glutamyl transferase in human serum and plasma. Siemens Material Number (SMN): ADVIA Chemistry: 10309495 (20 mL) 10316298 (70 mL)
Status
Open, Classified
Initiated
2020-11-23
Posted
not on file
Terminated
not on file
Reason
Gamma-Glutamyl Transferase (GGT) Reagent Low End Repeatability-Imprecision. Reagent may demonstrate a coefficient of variation (%CV) outside of the performance data for repeatability (within-run) listed in the Instructions For Use (IFU) on some analyzers.
Root cause
Device Design
Action
Siemens Healthcare Diagnostics Urgent Field Safety Notice letter beginning November 23, 2020 via Fed Ex and outside the US the distribution is determined at the country level. Letter states reason for recall , health risk and action to take: For the product listed above, please perform the following steps: 1. Review and apply the preliminary precision performance characteristics of d8% CV at a GGT concentration of approximately 27 U/L 42 U/L provided. 2. Please review this letter with your Medical Director. 3. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. 4. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Customer Care Center or your local Siemens technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Contact your Siemens Customer Care Center or your local Siemens technical support representative. The precision section of the IFUs will be updated when the investigation is complete. The updated IFUs will be uploaded into Document Library where all registered users who opt in to receive alerts will be notified of the updated IFU.
Quantity
397 units US; 1713 units OUS
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Australia, Austria, Bahamas, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Cura¿ao,¿St¿Eus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, India, Iran, Ireland, Israel, Italy, Japan, Kenya, Kuwait, Latvia, Libya, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, P.R. China, Pakistan, Philippines, Poland, Portugal, Qatar, Republic Korea, Romania, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, U.A.E., United Kingdom, Uruguay, Vatican City, Vietnam. ¿¿¿¿ ¿¿¿¿¿
Lot, serial or model codes
All lot codes. UDI: 10309495 (20 mL) = 00630414508344 10316298 (70 mL) = 00630414209753
510(k) numbers
not on file
PMA numbers
not on file
Product code
JQA
Firm FEI
2432235
Firm city
Tarrytown
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register