Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer Biomet, Inc.: Zimmer Segmental System, ZSS Distal Femur. Sizes as follows: a. Size B-LT (Item No. 00585001201); b. Size B-RT (Item No. 00585001202); c. Size C-LT (Item No. 00585001301); d. Size C-RT (Item No. 00585001302) Components of the system are intended to replace the proximal femur, mid-shaft femur, distal femur, proximal tibia, and/or total knee in cases that require extensive resection and restoration.

Recall number
Z-0721-2018
Recalling firm
Zimmer Biomet, Inc.
Product
Zimmer Segmental System, ZSS Distal Femur. Sizes as follows: a. Size B-LT (Item No. 00585001201); b. Size B-RT (Item No. 00585001202); c. Size C-LT (Item No. 00585001301); d. Size C-RT (Item No. 00585001302) Components of the system are intended to replace the proximal femur, mid-shaft femur, distal femur, proximal tibia, and/or total knee in cases that require extensive resection and restoration.
Status
Terminated
Initiated
2017-08-01
Posted
not on file
Terminated
2020-03-18
Reason
The products being recalled exceeded the weight previously tested leading to a possibility of damage to the package and/or a compromised sterile barrier.
Root cause
Package design/selection
Action
The firm, Zimmer Biomet, sent an "URGENT MEDICAL DEVICE RECALL REMOVAL"-Lot Specific letter dated 8/1/2017 to its Customers via email and/or FedEx. The letter describes the product, problem and actions to be taken. The Instructions to Distributors, Sales Representatives, and Distributor Operation Managers included to ensure affected team member are aware of the contents of the notification letter, immediately locate and quarantine affected product in inventory, complete and return the Certification of Acknowledgement form by email to: corporatequality.postmarket@zimmerbiomet.com., return all affected product, and note if there are any hospitals and/or surgeons for which need to be notified. Instructions for Risk Managers and surgeons are to ensure affected personnel are aware of the contents of the notification letter, quarantine any affected product on hand, complete and return the Certificate of Acknowledgement form by email to: corporatequality.postmarket@zimmerbiomet.com., and return any affected product on hand. For questions or concerns, please call the 411 call center at 1-574-371-3071 between 8:00 am and 5:00 pm EST, Monday through Friday or email corporatequality.postmarket@zimmerbiomet.com
Quantity
3,350 units total
Distribution
Worldwide Distribution: US (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, and Puerto Rico; and to countries of: Austria, Australia, Bahamas, Brazil, Canada, China, Dominican Republic, England, France, Germany, Israel, Japan, South Korea, Lebanon, Malaysia, Mexico, Netherlands, Panama, Singapore, Spain, Switzerland, and Thailand.
Lot, serial or model codes
a. Item No. 00585001201 (Lot Expiration Date Before 07/31/2026);  b. Item No. 00585001202 (Lot Expiration Date Before 07/31/2026);  c. Item No. 00585001301 (Lot Expiration Date Before 08/31/2026);  d. Item No. 00585001302 (Lot Expiration Date Before 07/31/2026)
510(k) numbers
["K070978"]
PMA numbers
not on file
Product code
KRO
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register