Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott Laboratories, Inc: AEROSET System Software version 1.00ER005 Catalog Number 2-89996-01 and 1.00ER005.2 Catalog Number 2-89996-02, distributed by Abbott Laboratories

Recall number
Z-0721-03
Recalling firm
Abbott Laboratories, Inc
Product
AEROSET System Software version 1.00ER005 Catalog Number 2-89996-01 and 1.00ER005.2 Catalog Number 2-89996-02, distributed by Abbott Laboratories
Status
Terminated
Initiated
2002-12-12
Posted
2003-04-15
Terminated
2004-10-28
Reason
The autoprint feature allows the switch of patient IDs in remote cases.
Root cause
Other
Action
The firm sent a Device Correction letter to all consignees on 12/12/2002, requesting that the users make adjustments to their systems, and retain the letter in their user manual. The letter included a customer reply form. The software will be corrected with the release of the next version.
Quantity
640
Distribution
Nationwide and to the following foreign countries: Mexico, Columbia, Dominican Republic, Hong Kong, Guatemala, Argentina, Canada, Australia, Venezuela, Brazil, Germany, New Zeland
Lot, serial or model codes
Version 1.00ER005 Catalog Number 2-89996-01 and Version 1.00ER005.2 Catalog Number 2-89996-02  AEROSET System software is not controlled by Control or Lot numbers.
510(k) numbers
not on file
PMA numbers
not on file
Product code
JJE
Firm FEI
1628664
Firm city
Irving
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this source