Mindray DS USA, Inc. dba Mindray North America: Sevoflurane (SEVO) V90 Electronic Vaporizer with the A9 Anesthesia System. Labeled as the following: 1. Sev E_Vap. (Safety Filling adapter), Part Number: 115-066756-00. 2. Exchange Sev E_Vap. (Safety Filling adapter), Part Number: E115-066756-00. Product Description: The Anesthesia System is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The A9 Anesthesia System utilizes V90 electronic vaporizers for anesthetic agent delivery.

Recall number
Z-0720-2026
Recalling firm
Mindray DS USA, Inc. dba Mindray North America
Product
Sevoflurane (SEVO) V90 Electronic Vaporizer with the A9 Anesthesia System. Labeled as the following: 1. Sev E_Vap. (Safety Filling adapter), Part Number: 115-066756-00. 2. Exchange Sev E_Vap. (Safety Filling adapter), Part Number: E115-066756-00. Product Description: The Anesthesia System is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The A9 Anesthesia System utilizes V90 electronic vaporizers for anesthetic agent delivery.
Status
Open, Classified
Initiated
2025-10-20
Posted
2025-11-26
Terminated
not on file
Reason
Potential for anesthesia leakage.
Root cause
Device Design
Action
Mindray notified consignees on about 10/20/2025 via letter. Consignees were informed that they could continue using the units as normal, but if an anesthetic odor or leak is detected, contact Mindray Technical Support. Interventions may include switching to another V90 vaporizer or using intravenous anesthesia as clinically appropriate. For vaporizer replacement procedures, refer to A9 Operator s Manual (Part No. H-046-017199-00, Section 3.4 Vaporizer ), available on the Mindray website. Mindray s Technical Support team can assist and may be reached at (877) 913-9663 (Option 2) Monday through Friday, 8:30 a.m. 5:30 p.m. ET. Consignees were instructed to complete and return the Acknowledgement and Receipt form provided, ensure that all applicable clinical staff are aware of the issue and maintain a copy of the recall notification with affected units or IFUs. Additionally, they were instructed to notify any organization to which the affected V90 electronic vaporizers identified have been distributed. The Mindray Care Team will contact consignees to arrange the replacement of all affected SEVO and ISO V90 electronic vaporizers located at their facility. The V90 electronic vaporizers will be shipped directly to each facility, and a Mindray Care Team representative will schedule the installation and collect the affected V90 electronic vaporizers.
Quantity
148 units
Distribution
Worldwide - US Nationwide distribution in the states of California, Florida, Georgia, Kentucky, Maryland, Mississippi, Missouri, New Jersey, Puerto Rico, Texas, Washington, West Virginia and the country of Canada.
Lot, serial or model codes
1. Part Number: 115-066756-00; DI: 06936415975777; Serial Numbers: AH6A45000871, AH6A45000880, AH6A45000881, AH6A46000929, AH6A46000931, AH6A46000944, AH6A47001200, AH6A44000657, AH6A44000658, AH6A44000661, AH6A44000664, AH6A44000677, AH6A44000678, AH6A44000680, AH6A44000768, AH6A44000778, AH6A46000925, AH6A46000927, AH6A46000928, AH6A46000935, AH6A46000937, AH6A46001114, AH6A47001159, AH6A47001164, AH6A47001174, AH6A47001201, AH6A45000872, AH6A46000932, AH6A46000939, AH6A46000940, AH6A47001196, AH6A47001202, AH6A47001207, AH6A47001215, AH6A46001102, AH6A46001111, AH6A46001116, AH6A47001160, AH6A47001161, AH6A47001163, AH6A47001165, AH6A47001166, AH6A47001167, AH6A47001178, AH6A46001097, AH6A46001117, AH6A44000645, AH6A44000647, AH6A44000648, AH6A44000651, AH6A44000655, AH6A44000656, AH6A44000659, AH6A44000660, AH6A44000662, AH6A44000684, AH6A44000686, AH6A44000687, AH6A44000688, AH6A44000690, AH6A44000691, AH6A44000695, AH6A44000696, AH6A44000698, AH6A44000700, AH6A44000702, AH6A44000703, AH6A44000770, AH6A44000776, AH6A46001090, AH6A46001109, AH6A46001118, AH6A47001171, AH6A47001176, AH6A44000771, AH6A44000772, AH6A45000815, AH6A45000819, AH6A45000821, AH6A45000822, AH6A45000824, AH6A45000825, AH6A45000829, AH6A45000832, AH6A45000834, AH6A45000835, AH6A45000868, AH6A45000869, AH6A45000870, AH6A45000873, AH6A45000874, AH6A45000875, AH6A45000876, AH6A45000877, AH6A45000878, AH6A45000882, AH6A45000883, AH6A45000884, AH6A46000926, AH6A46000930, AH6A46000936, AH6A46000943, AH6A44000779, AH6A44000674, AH6A44000692, AH6A44000777, AH6A45000879, AH6A47001170, AH6A44000765, AH6A44000769, AH6A47001175, AH6A28000535, AH6A46000918, AH6A46000923, AH6A46000924, AH6A46000934, AH6A46000938, AH6A46001091, AH6A46001093, AH6A46001099, AH6A46001100, AH6A46001101, AH6A46001103, AH6A46001104, AH6A46001107, AH6A46001108, AH6A46001110, AH6A46001113, AH6A47001177, AH6A44000679, AH6A44000683, AH6A44000689, AH6A44000697, AH6A44000649, AH6A44000650, AH6A44000652, AH6A44000676, AH6A44000681, AH6A44000682, AH6A44000699, AH6A44000704, AH6A44000766, AH6A44000773, AH6A44000775, AH6A44000780, AH6A47001216, AH6A44000693. 2. Part Number: E115-066756-00; DI: 06936415975777; Serial Number: AH6A45000885.
510(k) numbers
["K201957"]
PMA numbers
not on file
Product code
BSZ
Firm FEI
2221819
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register