Cordis US Corp: Cordis Angiographic Catheter Extensions. Used to transport fluid from the power injector to the catheter for injection into the patient.

Recall number
Z-0720-2023
Recalling firm
Cordis US Corp
Product
Cordis Angiographic Catheter Extensions. Used to transport fluid from the power injector to the catheter for injection into the patient.
Status
Open, Classified
Initiated
2022-11-07
Posted
2022-12-16
Terminated
not on file
Reason
There is a potential for separation at the male connector.
Root cause
Under Investigation by firm
Action
Cordis notified consignees on approximately 11/07/2022 via letter titled "Urgent MEDICAL DEVICE RECALL" sent to the Materials Director and Risk Manager. The letter instructed consignees to identify and segregate any affected product on hand, complete and return the Acknowledgement Form by email: GMB-CordisFieldAction@cordis.com, and return all affected product. Additionally, customers were instructed to share the letter with others in your facility who need to be made aware of this recall and with any other facility that may have been sent the affected units. OUS consignee notification will be conducted by Cordis affiliates in the affected countries / regions. For any health care professional with medical concerns, please contact Cordis to speak to a clinician at 1-800-327-7714 Option 1, Monday through Friday from 9:00 AM to 5:00 PM EDT. For questions related to assistance returning product or billing concerns please contact your sales representative or Cordis Customer Service at 1-800-327-7714 Option 1, Monday through Friday from 7:00 AM to 8:00 PM EDT. For questions related to the recall and/or acknowledgement form that are not adequately addressed in this letter, please contact Cordis QA at: GMB-CordisFieldAction@cordis.com.
Quantity
16,950 units (2,915 US and 14,000 OUS)
Distribution
Nationwide distribution to AZ, CA, DC, FL, GA, KY, LA, MD, MS, NY, OH, TX, WV, WY and PR. International distribution to Japan, Korea, Republic of, United Arab Emirates, Austria, Belgium, Czech Republic, Germany, Spain, France, United Kingdom, Iceland, Ireland, Israel, Netherlands, Portugal, Russia.
Lot, serial or model codes
UDI-DI: (01)2070532064702; Catalog Number (Lot Number): 502-100D (18068258, 18069744, 18101041, 18111609); 502-101D (18062881, 18081307, 18064731, 18086388, 18066535, 18097369, 18070824, 18102623, 18077497); 502-102D (18060997, 18078225, 18099118, 18062882, 18079156, 18104313, 18064732, 18083133, 18106093, 18072508, 18084892, 18108019, 18074058, 18088193, 18109512, 18076420, 18090247)
510(k) numbers
K970854
PMA numbers
not on file
Product code
DQO
Firm FEI
1016427
Firm city
Miami Lakes
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this source