FDA Medical Device Recalls (openFDA)
Philips North America: Incisive CT, Model Numbers 728144 and 728143. Computed Tomography X-Ray System PIM cable part number 459801179871
- Recall number
- Z-0719-2024
- Recalling firm
- Philips North America
- Product
- Incisive CT, Model Numbers 728144 and 728143. Computed Tomography X-Ray System PIM cable part number 459801179871
- Status
- Open, Classified
- Initiated
- 2023-12-07
- Posted
- 2024-01-12
- Terminated
- not on file
- Reason
- If the Patient Interface Monitor (PIM) Cable is plugged/unplugged frequently as part of daily operation, the connector and wire may disconnect from the welding point, which may result in intermittent loss of the ECG signal. Loss of ECG signal may lead to the decision to rescan the patient or delay in diagnosis during a clinical emergency.
- Root cause
- Component design/selection
- Action
- On December 7, 2023, the firm notified affected customers through URGENT Medical Device Correction Notice letters. Customers were informed that they may continue to use their Philips Incisive CT system in accordance with the intended use. However, to minimize the potential for loss of the ECG signal, customers should avoid frequently plugging and unplugging the PIM cable. In case of a connection failure, contact your local service representative and reference FCO72800782. Philips will contact you to schedule time for a Field Service Engineer to visit your site to replace the PIM cable with an upgraded version. If you need additional information or support concerning this issue, please contact Philips' Customer Care Solutions Center at 1-800-722-9377.
- Quantity
- 55 US; 1029 ROW
- Distribution
- Domestic distribution to the following states: AL CA CO FL GA HI IL IN KS KY LA MD MI MO MS ND NE NJ NY OH OK PA RI SC TX UT VT WA International distribution worldwide.
- Lot, serial or model codes
- Incisive CT systems configured with the PIM cable option are affected by this issue. To determine whether your system is affected, please locate the connection cable between your ECG monitor and incisive CT gantry and verify that the part number on your PIM cable matches 459801179871 . (1) REF 728143 - UDI-DI 00884838085015 Serial Numbers: 33007 33038 33013 33006 33012 33045 33011 500102 33015 33005 33041 33047 500217 500255 33035 500252 500145 500046 500168 33009 33046 33024 500222 500064 33034 500242 500301 33010 33019 33001 33020 33031 33037 500110 500111 500178 33003 500282 33002 33033 (2) REF 728144 - UDI-DI 00884838105508 Serial Numbers: 34002 34051 34050 34041 34059 34057 34054 34040 34007 34026 34058 34049 34056 34052
- 510(k) numbers
- ["K180015","K212441"]
- PMA numbers
- not on file
- Product code
- JAK
- Firm FEI
- 3006648320
- Firm city
- Cambridge
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28