FDA Medical Device Recalls (openFDA)
Philips Electronics North America Corporation: INTEGRIS BN/BV3000; Model Number: 72240 The Philips INTEGRIS 3000 I 5000 system is intended for interventional and diagnostic vascular and neurovascular procedures. The Philips INTEGRIS Allura 9 Biplane system is intended for diagnostic cardiovascular, vascular and interventional procedures. The Philips INTEGRIS Allura 12 & 15 system is intended for peripheral, abdominal, cerebral diagnostic and interventional angiography, neuro applications, cardiac applications and non-vascular interventions
- Recall number
- Z-0719-2016
- Recalling firm
- Philips Electronics North America Corporation
- Product
- INTEGRIS BN/BV3000; Model Number: 72240 The Philips INTEGRIS 3000 I 5000 system is intended for interventional and diagnostic vascular and neurovascular procedures. The Philips INTEGRIS Allura 9 Biplane system is intended for diagnostic cardiovascular, vascular and interventional procedures. The Philips INTEGRIS Allura 12 & 15 system is intended for peripheral, abdominal, cerebral diagnostic and interventional angiography, neuro applications, cardiac applications and non-vascular interventions
- Status
- Terminated
- Initiated
- 2015-11-03
- Posted
- 2016-01-30
- Terminated
- 2020-09-15
- Reason
- The use of Allura Xper, Allura Clarity, Allura CV, Allura Centron and UNIQ systems regarding live images and still images can lead to still images being interpreted as live Images.
- Root cause
- Device Design
- Action
- The firm, Philips, issued "URGENT-Field Safety Notices" dated 10/26/2015 and 11/19/2015 to its customers. The notices described the product, problem and actions to be taken. The customers were instructed to ensure the audible signal (X-ray buzzer) is always enabled, to activate only ONE fluoroscopy pedal at time, and to configure the footswitch to the default Philips configuration. Customers were also instructed to review the information with all members of your staff who need to be aware; retain a copy with the equipment IFU and complete and return the Confirmation Respond form to Mail Stop 4302, 3000 Minuteman Road, Andover, MA 01810 or email: Debra.Gibbs@Philips.com. Philips is tracking the delivery of these notices via certified return receipt. If you need any further information or support concerning this issue, please contact your local Philips representative: Technical Support Line: 1-800-722-9377.
- Quantity
- 13297 in total
- Distribution
- Worldwide Distribution: US (Nationwide) including Puerto Rico and countries of: Albania, Algeria, US Virgin Islands, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Democratic Congo, Costa Rica, Croatia, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Fiji, Finland, France, French Polynesia, Germany, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Martinique, Mauritius, Mexico, Moldova, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, and Vietnam.
- Lot, serial or model codes
- All serial numbers
- 510(k) numbers
- ["K971365"]
- PMA numbers
- not on file
- Product code
- IZI
- Firm FEI
- 1218950
- Firm city
- Andover
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28