FDA Medical Device Recalls (openFDA)
Siemens Healthcare Diagnostics, Inc.: IMMULITE 2000 Systems 3gAllergy Specific IgE Universal Kit Tests OUS, except Japan Siemens Material Number (SMN): 10380875 For the quantitative measurement of allergen-specific IgE in human serum, as an aid in the clinical diagnosis of IgE mediated allergic disorders.
- Recall number
- Z-0717-2023
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Product
- IMMULITE 2000 Systems 3gAllergy Specific IgE Universal Kit Tests OUS, except Japan Siemens Material Number (SMN): 10380875 For the quantitative measurement of allergen-specific IgE in human serum, as an aid in the clinical diagnosis of IgE mediated allergic disorders.
- Status
- Open, Classified
- Initiated
- 2022-11-04
- Posted
- 2022-12-15
- Terminated
- not on file
- Reason
- Potential for falsely elevated specific IgE mold allergen reactivity with quality control material and patient samples, may result in false reactive result
- Root cause
- Under Investigation by firm
- Action
- Siemens initiated issued on 11/04/2022 an Urgent Medical Device Correction (UMDC) IMC 23-01.A.US to US customers, an Urgent Field Safety Notice (UFSN) IMC 23-01.A-1.OUS to Outside the US (OUS, except Japan) customers, and an Urgent Field Safety Notice (UFSN) IMC 23-01.A-2.OUS (Japan) to customers in Japan . Letter states reason for recall, health risk and action to take: Please review this letter with your Medical Director. Discontinue use of the kit lots listed in Table 1. Review your inventory of this product to determine your laboratory s replacement needs and to provide information to Siemens for reporting to the authorities. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the product listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative
- Quantity
- 5,193 kits
- Distribution
- Worldwide distribution. US Nationwide, Argentina¿¿¿¿¿¿ Australia¿¿¿¿¿¿ Austria¿¿¿¿¿¿¿¿ Azerbaijan¿¿¿¿¿ Belgium¿¿¿¿¿¿¿¿ Brazil¿¿¿¿¿¿¿¿¿ Chile¿¿¿¿¿¿¿¿¿¿ Colombia¿¿¿¿¿¿¿ Costa¿Rica¿¿¿¿¿ Croatia¿¿¿¿¿¿¿¿ Czech¿Republic¿ Denmark¿¿¿¿¿¿¿¿ Ecuador¿¿¿¿¿¿¿¿ Egypt¿¿¿¿¿¿¿¿¿¿ Estonia¿¿¿¿¿¿¿¿ France¿¿¿¿¿¿¿¿¿ Germany¿¿¿¿¿¿¿¿ Guatemala¿¿¿¿¿¿ Hong¿Kong¿¿¿¿¿¿ Hungary¿¿¿¿¿¿¿¿ Indonesia¿¿¿¿¿¿ Israel¿¿¿¿¿¿¿¿¿ Italy¿¿ Japan¿¿¿¿¿¿¿¿ Latvia¿¿¿¿¿¿¿¿¿ Lithuania¿¿¿¿¿¿ Macedonia¿¿¿¿¿¿ Mexico¿¿¿¿¿¿¿¿¿ Netherlands¿¿¿¿ Norway¿¿¿¿¿¿¿¿¿ Pakistan¿¿¿¿¿¿¿ Panama¿¿¿¿¿¿¿¿¿ Paraguay¿¿¿¿¿¿¿ Philippines¿¿¿¿ Poland¿¿¿¿¿¿¿¿¿ Portugal¿¿¿¿¿¿¿ Romania¿¿¿¿¿¿¿¿ Russian¿Fed.¿¿¿ Serbia¿¿¿¿¿¿¿¿¿ Slovakia¿¿¿¿¿¿¿ Slovenia¿¿¿¿¿¿¿ South¿Africa¿¿¿ Spain¿¿¿¿¿¿¿¿¿¿ Sweden¿¿¿¿¿¿¿¿¿ Switzerland¿¿¿¿ Taiwan¿¿¿¿¿¿¿¿¿ Turkey¿¿¿¿¿¿¿¿¿ Ukraine¿¿¿¿¿¿¿¿ United¿Kingdom¿ Uruguay, and Uzbekistan¿¿¿¿¿
- Lot, serial or model codes
- UDI:-DI: (01)00630414962269(10)826(17)20221130; (01)00630414962269(10)830(17)20221130; (01)00630414962269(10)831(17)20221130; (01)00630414962269(10)832(17)20221130; (01)00630414962269(10)833(17)20221130. Kit Lots: 826, 830, 831,832, 833
- 510(k) numbers
- ["K101572"]
- PMA numbers
- not on file
- Product code
- DHB
- Firm FEI
- 1219913
- Firm city
- East Walpole
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28