B Braun Medical, Inc.: APEX Compounding System Control Panel Module, REF: 601224, Product Code AX1000 - Product Usage: intended for intravascular administration to a patient.

Recall number
Z-0717-2021
Recalling firm
B Braun Medical, Inc.
Product
APEX Compounding System Control Panel Module, REF: 601224, Product Code AX1000 - Product Usage: intended for intravascular administration to a patient.
Status
Terminated
Initiated
2020-11-23
Posted
not on file
Terminated
2023-03-14
Reason
There is the potential for the compounding system to not immediately interrupt compounding and alert user to the presence of an air bubble exceeding 3% of the ordered volume for an ingredient. This could result in a a compounded drug that does not meet the specified accuracy range for an individual ingredient.
Root cause
Software change control
Action
On Nov. 23, 2020, the firm issued a "Urgent Medical Device Correction" letter to affected consignees via USPS mail. In addition to informing consignees about the recall device, the firm requested consignees take the following actions: 1. Review the URGENT MEDICAL DEVICE CORRECTION Notification in its entirety and ensure that all users in your organization and other concerned persons are informed about this voluntary Medical Device Correction and the affected product(s). 2. Ensure workflow(s) are in place to review air bubbles and scrap any final containers which exceed your facilitys guidelines until your APEX Compounding System has been upgraded to Software Version 1.0.5.58 or higher. " When compounding is complete, if bubbles were detected, APEX activates the Bubble Events button on the Compounding Complete screen allowing you to review the bubbles detected during compounding. " Until such time as your APEX software has been updated, BBMI recommends having qualified pharmacy personnel review this screen prior to removing the final container to ensure that the size of any air bubbles are within an allowable range per your facilitys guidelines. " Alternatively, your facility can have qualified personnel review each Compounding Activity Report generated by the device which also contains air bubble information. " Instructions for viewing the Bubble Events screen are provided in Attachment 02 of this Notice. Instructions for printing and reviewing the Compounding Activity Report are provided in Attachment 03 of this Notice. " Any final containers which exceed your facilitys allowable range should be scrapped. 3. Return the completed Acknowledgement Medical Device Correction form to BBMI by utilizing the enclosed self-addressed envelope or via scanned e-mail to BBMUS_Carrollton_QualitySystemManagement@bbraunusa.com within two (2) weeks of receipt, even if the total inventory in your possession is zero (0). It is important this form is
Quantity
248 units
Distribution
US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, IA, IL, LA, MA, MD, MI, MN, MO, NC, NH, NJ, NV, NY, OH, OK, PA, TN,TX, WA.
Lot, serial or model codes
All APEX Compounding System with Software Version 1.0.5.57 or earlier
510(k) numbers
not on file
PMA numbers
not on file
Product code
NEP
Firm FEI
1641965
Firm city
Carrollton
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register