FDA Medical Device Recalls (openFDA)
Philips Electronics North America Corporation: Allura Xper systems with Certeray X-ray Generator located in regions with high mains (480V) voltage. Product Usage: The Allura Xper series is intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placements, embolisations and thrombolysis. Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures (such as PTCA, stent placing, atherectomies), pacemaker implanatations, and electrophysiology (EP). Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures.
- Recall number
- Z-0716-2018
- Recalling firm
- Philips Electronics North America Corporation
- Product
- Allura Xper systems with Certeray X-ray Generator located in regions with high mains (480V) voltage. Product Usage: The Allura Xper series is intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placements, embolisations and thrombolysis. Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures (such as PTCA, stent placing, atherectomies), pacemaker implanatations, and electrophysiology (EP). Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures.
- Status
- Terminated
- Initiated
- 2018-01-04
- Posted
- 2018-02-24
- Terminated
- 2019-10-09
- Reason
- There is an increase in the failure rate of certain Anode Drive Units (ADU5) used in these products. Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs (insulated-gate bipolar transistor) of the ADU. The overheating may also generate a peculiar burning smell that may be noticed. In some instances this burning smell may be noticed before the ADU fails. When the ADU fails the X-ray performance of the system is reduced to Emergency Fluoroscopy. Exposure will not be possible and image quality is reduced.
- Root cause
- Device Design
- Action
- Customer Notification letter dated August 2017 were sent to affected customers. The letter identified affected product, problem and actions to be taken. The customer is requested to place this Field Safety Notice with the documentation of the system until this correction has been implemented. For questions contact your local Philips representative.
- Quantity
- 705
- Distribution
- Worldwide Distribution - US Nationwide; foreign distribution to Brazil, Germany, Iceland, Japan, Mexico, and (updated 6/25/18) Saudi Arabia.
- Lot, serial or model codes
- Allura Xper systems - Model #'s 722026, 722027, 722028, 722029, 722033, 722034, 722035, 722036, 722038, 722039, 722058, 722059, 722063, 722064, 722065, 722066, 722078, 722079, (added as of 6/25/2018) 1120
- 510(k) numbers
- ["K133292","K141979"]
- PMA numbers
- not on file
- Product code
- OWB
- Firm FEI
- 1218950
- Firm city
- Andover
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28