Skip to the content

FDA Medical Device Recalls (openFDA)

Mobius Imaging, LLC: AIRO Mobile CT System, Model Number: MobiCT-32 Product Usage: The AIRO is intended to be used for X-ray computed tomography applications for anatomy that can be imaged in the 107cm aperture excluding pediatric patients and patients weighing over 400 lbs. (182 kg).

Recall number
Z-0715-2018
Recalling firm
Mobius Imaging, LLC
Product
AIRO Mobile CT System, Model Number: MobiCT-32 Product Usage: The AIRO is intended to be used for X-ray computed tomography applications for anatomy that can be imaged in the 107cm aperture excluding pediatric patients and patients weighing over 400 lbs. (182 kg).
Status
Terminated
Initiated
2017-09-26
Posted
not on file
Terminated
2018-08-16
Reason
If the tilt-drive motor has an intermittent connection or broken encoder or sensor wire there is a risk of unintended or unpredictable device motion while the CT system is homing the axis at startup or preparing for a scan.
Root cause
Under Investigation by firm
Action
Mobius sent an Urgent Field Safety Notice letter dated September 26, 2017 to their customers. The letter identified the affected product, problem and actions to be taken. Mobius Imaging is advising all customers to discontinue using the AIRO until update is completed on their system. Please provide a copy of this Field Safety Notification to all AIRO operators that use the AIRO system, and quarantine the system per hospital policies and procedures.
Quantity
105
Distribution
Worldwide Distribution - US Nationwide in the states of: AZ, CA, CO, FL, GA, IL, KY, LA, MA, MD, MI, MN, MO, MT, NC, NY, OH, OR, PA, RI, TN, TX, UT, VA, WA, WV and the countries of Argentina, Australia, Austria, Belgium, Canada, Egypt, France, Germany, India, Italy, Japan, Kazakhstan, Lithuania, Netherlands, Pakistan, Russia, Saudi Arabia, Spain, Switzerland, UAE, Uzebekistan
Lot, serial or model codes
Requested.
510(k) numbers
["K160126"]
PMA numbers
not on file
Product code
JAK
Firm FEI
3010151377
Firm city
Shirley
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register