FDA Medical Device Recalls (openFDA)
Mobius Imaging, LLC: AIRO Mobile CT System, Model Number: MobiCT-32 Product Usage: The AIRO is intended to be used for X-ray computed tomography applications for anatomy that can be imaged in the 107cm aperture excluding pediatric patients and patients weighing over 400 lbs. (182 kg).
- Recall number
- Z-0715-2018
- Recalling firm
- Mobius Imaging, LLC
- Product
- AIRO Mobile CT System, Model Number: MobiCT-32 Product Usage: The AIRO is intended to be used for X-ray computed tomography applications for anatomy that can be imaged in the 107cm aperture excluding pediatric patients and patients weighing over 400 lbs. (182 kg).
- Status
- Terminated
- Initiated
- 2017-09-26
- Posted
- not on file
- Terminated
- 2018-08-16
- Reason
- If the tilt-drive motor has an intermittent connection or broken encoder or sensor wire there is a risk of unintended or unpredictable device motion while the CT system is homing the axis at startup or preparing for a scan.
- Root cause
- Under Investigation by firm
- Action
- Mobius sent an Urgent Field Safety Notice letter dated September 26, 2017 to their customers. The letter identified the affected product, problem and actions to be taken. Mobius Imaging is advising all customers to discontinue using the AIRO until update is completed on their system. Please provide a copy of this Field Safety Notification to all AIRO operators that use the AIRO system, and quarantine the system per hospital policies and procedures.
- Quantity
- 105
- Distribution
- Worldwide Distribution - US Nationwide in the states of: AZ, CA, CO, FL, GA, IL, KY, LA, MA, MD, MI, MN, MO, MT, NC, NY, OH, OR, PA, RI, TN, TX, UT, VA, WA, WV and the countries of Argentina, Australia, Austria, Belgium, Canada, Egypt, France, Germany, India, Italy, Japan, Kazakhstan, Lithuania, Netherlands, Pakistan, Russia, Saudi Arabia, Spain, Switzerland, UAE, Uzebekistan
- Lot, serial or model codes
- Requested.
- 510(k) numbers
- ["K160126"]
- PMA numbers
- not on file
- Product code
- JAK
- Firm FEI
- 3010151377
- Firm city
- Shirley
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28