Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corporation: Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 6) - Product Usage: intended to be used for the controlled administration of fluids.

Recall number
Z-0713-2021
Recalling firm
Baxter Healthcare Corporation
Product
Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 6) - Product Usage: intended to be used for the controlled administration of fluids.
Status
Open, Classified
Initiated
2020-11-23
Posted
not on file
Terminated
not on file
Reason
There is a potential software error during programming.
Root cause
Under Investigation by firm
Action
A Medical Device Correction communication will be sent to all affected customers via U.S.P.S., first class mail. All pumps will be updated with new software which will prevent the malfunction from occurring. The software upgrade will be implemented upon next service or a local Baxter service representative will contact the facility to determine the correction plan and schedule the software update.
Quantity
293,926 devices
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Canada, Barbados, Jamaica, Bermuda, Guyana, Bahamas, and Trinidad and Tobago.
Lot, serial or model codes
Product Code 35700BAX, GTIN 00085412091587
510(k) numbers
["K042121"]
PMA numbers
not on file
Product code
FRN
Firm FEI
1417572
Firm city
Deerfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register