Radiometer America Inc: pO2 D999 membrane units. Distributed by Radiometer America, Inc., Westlake, OH 44145 pO2 membrane units, ordering number 942-042 Intended use is in vitro testing of samples of whole blood for pO2 (partial pressure of oxygen).
- Recall number
- Z-0713-2014
- Recalling firm
- Radiometer America Inc
- Product
- pO2 D999 membrane units. Distributed by Radiometer America, Inc., Westlake, OH 44145 pO2 membrane units, ordering number 942-042 Intended use is in vitro testing of samples of whole blood for pO2 (partial pressure of oxygen).
- Status
- Terminated
- Initiated
- 2013-06-10
- Posted
- 2014-01-10
- Terminated
- 2014-06-19
- Reason
- RADIOMETER became aware that the regarded lot pO2 units 942-042R0971 were produced with inner solution for pCO2 membrane units. This solution is a salt solution similar to the pO2 inner solution, but also contains glycerol which can be reduced at the cathode causing a suddenly rising zero current, causing the membrane unit to fail the following Cal2 calibration.
- Root cause
- Mixed-up of materials/components
- Action
- Radiometer sent a Product Notification letter dated June 18, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to check their inventory and remove any affected product, discard all affected product in their possession, and complete page 2 of the letter and return to their Radiometer representative. Radiometer will replace any quantity of affected product they have at their location. Customers with questions were instructed to contact Radiometer America Technicial Support at 1-800-736-0600, opt 4. For questions regarding this recall call 440-871-8900, ext 209.
- Quantity
- 209 boxes/6 membranes
- Distribution
- Worldwide Distribution - US (nationwide) including states of: CO, CT, GA, MA, MO & MS and country of: Canada.
- Lot, serial or model codes
- Part #: 942-042, Lot #: R0971
- 510(k) numbers
- ["K895181","K922922","K935655"]
- PMA numbers
- not on file
- Product code
- CHL
- Firm FEI
- 1523456
- Firm city
- Westlake
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28