Skip to the content

FDA Medical Device Recalls (openFDA)

Atrium Medical Corporation: ProLite Mesh - mesh, surgical, polymeric ProLite Mesh is intended for use in soft tissue deficiencies including hernia repair, traumatic or surgical wounds and chest wall reconstruction procedures requiring reinforcement with a non-absorbable supportive material.

Recall number
Z-0712-2018
Recalling firm
Atrium Medical Corporation
Product
ProLite Mesh - mesh, surgical, polymeric ProLite Mesh is intended for use in soft tissue deficiencies including hernia repair, traumatic or surgical wounds and chest wall reconstruction procedures requiring reinforcement with a non-absorbable supportive material.
Status
Terminated
Initiated
2017-11-22
Posted
2018-02-23
Terminated
2018-07-30
Reason
An incorrect ProLite Mesh configuration was packaged as product code 1010306-06, with lot number 412298. ProLite Mesh manufactured as product code 1010306-04 was inadvertently placed inside the pouch of packages labeled with product code 1010306-06 and lot number 412298.
Root cause
Under Investigation by firm
Action
Please examine inventory to determine if you have any of the affected products listed on page 1. If so, please remove affected products, quarantine them, and place in a secure location. If you have affected product, please contact Getinge Customer Service at internationalcs@atriummed.com for a Return Authorization (RA) and shipping instructions to return any affected product. Pack the product to be returned with the appropriate return documents, and using the shipping instructions provided, arrange for pick up with the designated delivery service provider. Please complete and sign the enclosed Urgent Medical Device Field Safety Notice to acknowledge receipt. Return the completed form to the following e-mail address: MESH2017@getinge.com For further questions, please call (973) 709-7654.
Quantity
60 units
Distribution
Internationally to Spain only.
Lot, serial or model codes
Product Code/REF 1010306-06 Lot Number/LOT: 412298
510(k) numbers
["K151437"]
PMA numbers
not on file
Product code
FTL
Firm FEI
3011175548
Firm city
Merrimack
Firm state
NH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register