Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA Inc: ECAT Emerge PET Scanners (distributed by Siemens Medical Solutions USA, Inc.), ECAT ART PET Scanners (distributed by Siemens Medical Solutions USA, Inc.) and Sceptre PET Scanners (distributed by Hitachi Medical Systems of America)

Recall number
Z-0712-06
Recalling firm
Siemens Medical Solutions USA Inc
Product
ECAT Emerge PET Scanners (distributed by Siemens Medical Solutions USA, Inc.), ECAT ART PET Scanners (distributed by Siemens Medical Solutions USA, Inc.) and Sceptre PET Scanners (distributed by Hitachi Medical Systems of America)
Status
Terminated
Initiated
2006-03-28
Posted
2006-04-04
Terminated
2009-12-22
Reason
There is a potential for inaccurate, elevated SUVs (standardized uptake values). On follow-up scans, the elevated SUV could be misinterpreted, and inappropriately impact a patient's treatment plan.
Root cause
Other
Action
Written notification was sent out beginning 03/28/2006 to end users to advise them of the problem. The notification also included specific work instructions to aid the end user until the patch is available and the upgrade could be performed by a service representative.
Quantity
57 devices
Distribution
KS, WA, CA, OH, FL, TX, Germany, Great Britain, Canada Switzerland, Turkey, Argentina, Austria, France, & Spain
Lot, serial or model codes
software versions 7.2.2 rel 2, 7.2.2 rel 4A, 7.2.2 rel 5 and 7.3 Model: ART Model Number: 3600066-00 Serial numbers: 1001, 1002, 1003, 1004, 1005, 1006, 1007, 1008, 1009, 1010, 1012, 1013, 1014  Model: ART Model Number: 3600078-00 Serial numbers: 1001, 1002, 1003  Model: ART Model Number: 3600082-00 Serial numbers: 1002  Model: ART Model Number: 3600084-00 Serial numbers: 1001, 1002, 1003, 1004, 1005, 1006,   Model: ART 2000 Model Number: 3600090-00 Serial numbers: 1001, 1002, 1003, 1004, 201005, 201006  Model: EMERGE, EMERGE (Toshiba), EMERGE (Sceptre - Hitachi), EMERGE m (Sceptre - Hitachi), PICO EMERGE (Sceptre-Hitachi), PICO EMERGEm (Sceptre - Hitachi) Model Number: 3600099-00 Serial numbers: 101001, 201002, 201003, 201004, 201005, 201006, 201007, 201008, 201009, 201010, 201011, 201012, 301013, 301014, 301015, 301016, 301017, 301018, 301019, 301020, 301022, 301023, 301024, 301025, 301026, 301027, 301028  Model Number: PICO EMERGE (Sceptre-Hitachi) Model Number: P008060 Serial number: 401001
510(k) numbers
["K003241","K940478"]
PMA numbers
not on file
Product code
KPS
Firm FEI
1034973
Firm city
Knoxville
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register