FDA Medical Device Recalls (openFDA)
Maquet Cardiovascular, LLC: Product REF 4050-100N, DRAIN, EXPRESS BRU W/AC, Blood Recovery Unit
- Recall number
- Z-0710-2024
- Recalling firm
- Maquet Cardiovascular, LLC
- Product
- Product REF 4050-100N, DRAIN, EXPRESS BRU W/AC, Blood Recovery Unit
- Status
- Open, Classified
- Initiated
- 2023-12-13
- Posted
- 2024-01-12
- Terminated
- not on file
- Reason
- Express chest drains are pre-packaged with sterile water syringes which have bene recalled. If the device is not kept in an upright position, the patient may be exposed to infectious material. Subsequent hazards include but are not limited to infection, abscess, sepsis, and death.
- Root cause
- Material/Component Contamination
- Action
- On December 13, 2023, the firm notified customers through "URGENT MEDICAL DEVICE CORRECTION" letters. Customers were instructed to examine their inventory to determine if they had any affected lots of the Atrium Express Dry Suction Dry Seal Chest Drains. The correction only affects the sterile water syringe; do not use this syringe. If customers do not need visualization of active pneumothorax, they may keep the affected Express chest drain and set up without water. If customers wish to use the Express chest drain as intended, they may replace the affected sterile water syringe with a new syringe filled with sterile water using aseptic technique. Customers also have the option of returning affected Express chest drains to the firm for credit. If you have any questions, please contact your Getinge representative or call the Getinge Customer Support at (888) 943-8872 (press option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m.(Eastern Time Zone).
- Quantity
- (7,501 cases/45,006 devices US); no OUS
- Distribution
- Worldwide distribution: US (nationwide) and OUS (Foreign) to countries of: Australia, Brazil, Canada, Luxembourg, Netherlands, Norway, Spain and UK
- Lot, serial or model codes
- UDI-DI 00650862115147 Lots 499345 498062 496693 495731 494226 492645 492078 490763 487811 490139 489162 487849 483249 470644 468860 467477 466952
- 510(k) numbers
- ["K201305"]
- PMA numbers
- not on file
- Product code
- CAC
- Firm FEI
- 2242352
- Firm city
- Wayne
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28