St Jude Medical: Implantable Cardioverter-Defibrillator Models affected include the following: Atlas+ Models V-193, V-193C and V-243 Epic+ Models V-196, V-236 and V-239 Epic Models V-197 and V-233 Atlas Model V-242 Epic HF Models V-337 and V-338 Epic+HF Models V-339 and V-350 Atlas+ HF Models V-340, V-341 and V-343

Recall number
Z-0709-05
Recalling firm
St Jude Medical
Product
Implantable Cardioverter-Defibrillator Models affected include the following: Atlas+ Models V-193, V-193C and V-243 Epic+ Models V-196, V-236 and V-239 Epic Models V-197 and V-233 Atlas Model V-242 Epic HF Models V-337 and V-338 Epic+HF Models V-339 and V-350 Atlas+ HF Models V-340, V-341 and V-343
Status
Terminated
Initiated
2005-03-10
Posted
2005-04-13
Terminated
2005-07-11
Reason
Sensitivity of a magnetic switch was not being set correctly. This could result in delivery of an unneeded shock during surgery with use of electrocautery.
Root cause
Other
Action
Notification materials were distributed to sales representatives on March 10, 2005. They will visit each doctor that is following the devices.
Quantity
1488
Distribution
Nationwide, Canada, Australia, Sweden, Brazil, Colombia, Argentina.
Lot, serial or model codes
Too numerous to include here.
510(k) numbers
not on file
PMA numbers
["P910023","P030054"]
Product code
NIK
Firm FEI
2017865
Firm city
Sylmar
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register