Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Axiom Artis, Model Numbers: 5904433, 5904441, 5904649, 5904656, 5917054, 7007755, 7008605, 7412807, 7413078, 7555357, 7555373, 7727717, 7728350 & 7728392

Recall number
Z-0708-2018
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Axiom Artis, Model Numbers: 5904433, 5904441, 5904649, 5904656, 5917054, 7007755, 7008605, 7412807, 7413078, 7555357, 7555373, 7727717, 7728350 & 7728392
Status
Terminated
Initiated
2017-06-06
Posted
not on file
Terminated
2019-05-16
Reason
Potential for liquid entry
Root cause
Device Design
Action
On June 6, 2017, Siemens Healthineers distributed URGENT MEDICAL DEVICE CORRECTION notices to their customers via USPS Certified Mail. Siemens conducted a corrective field action in June 2015 to protect the Artis table against liquid entry at critical locations. However, even with this corrective action, if there are larger amounts of liquid on the table surface, they can still enter the table. Siemens Medical Solutions USA, Inc. will contact customers to arrange a date to perform the corrective action. Please feel free to contact the service organization for an earlier appointment by calling 610-448-6461.
Quantity
not on file
Distribution
U.S. Nationwide
Lot, serial or model codes
10094135, 10094137, 10094141, 10094142, 10094143, 10280959, 10848280, 10848281, 10848282, 10848353, 10848354, 10848355, 5904433, 5904441, 5904649, 5904656, 5917054, 7007755, 7008605, 7412807, 7413078, 7555357, 7555373, 7727717, 7728350, 7728392
510(k) numbers
["K052202"]
PMA numbers
not on file
Product code
OWB
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register