Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Navigation, Inc.: Medtronic Adapter Percutaneous Pin, REF 9734752 (GTIN 00643169529489) and REF 9734752-G02 (GTIN 00763000054915), contained in the following kits: Upgrade Set Percutaneous Pin, REF 9735502 (GTIN 00643169529533) and REF 9735502-G02 (GTIN 00763000208462); Medtronic Adapter Percutaneous Pin Kit, REF 9734752K (GTIN 00643169540521); and Medtronic Adapter Percutaneous Pin Kit, REF 9734752K (GTIN 00643169641006) (Japan). The percutaneous pin adapter is a component of the StealthAir Frame Assembly.

Recall number
Z-0707-2020
Recalling firm
Medtronic Navigation, Inc.
Product
Medtronic Adapter Percutaneous Pin, REF 9734752 (GTIN 00643169529489) and REF 9734752-G02 (GTIN 00763000054915), contained in the following kits: Upgrade Set Percutaneous Pin, REF 9735502 (GTIN 00643169529533) and REF 9735502-G02 (GTIN 00763000208462); Medtronic Adapter Percutaneous Pin Kit, REF 9734752K (GTIN 00643169540521); and Medtronic Adapter Percutaneous Pin Kit, REF 9734752K (GTIN 00643169641006) (Japan). The percutaneous pin adapter is a component of the StealthAir Frame Assembly.
Status
Terminated
Initiated
2018-12-06
Posted
not on file
Terminated
2020-12-17
Reason
Under certain circumstances, the percutaneous pin adapter used with the StealthAir Frame Assembly may slightly rotate after surgical placement and not return to its original position, even when connections are tight and secure.
Root cause
Device Design
Action
The recalling firm issued a Dear Healthcare Professional letter dated 12/6/2018 via FedEx explaining the issue and requesting the consignee examine their inventory and quarantine the percutaneous pin adapter for return to the firm. Only the percutaneous pin adapter was to be returned and not the remaining items in the assembly.
Quantity
1,297 percutaneous pin adapters
Distribution
Distribution was nationwide. There was government/military/foreign distribution.
Lot, serial or model codes
Percutaneous pin adapter lot numbers: 150114, 150505, 150506, 150904, 151005, 151007 160201, 160224, 160419, 160617, 160707, 170216, 180110, and 180409.
510(k) numbers
["K131425"]
PMA numbers
not on file
Product code
OLO
Firm FEI
1000517638
Firm city
Louisville
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register