FDA Medical Device Recalls (openFDA)
Ev3, Inc: ev3 Protege GPS Self-Expanding Nitinol Stent Biliary System--- Sterilization with Ethylene Oxide Gas. ev3, 4600 Nathan Lane North, Plymouth, MN 55442-2920 USA. The self-expanding stent is made of nickel titanium alloy (Nitinol) and comes pre-mounted on an over-the-wire delivery system. The stent is cut from a nitinol tube into an open lattice design and is designed with tantalum radiopaque markers at the proximal and distal ends of the stent. Upon deployment, the stent achieves its predetermined diameter and exerts a constant, gentle outward force to establish patency in the biliary ducts.
- Recall number
- Z-0707-06
- Recalling firm
- Ev3, Inc
- Product
- ev3 Protege GPS Self-Expanding Nitinol Stent Biliary System--- Sterilization with Ethylene Oxide Gas. ev3, 4600 Nathan Lane North, Plymouth, MN 55442-2920 USA. The self-expanding stent is made of nickel titanium alloy (Nitinol) and comes pre-mounted on an over-the-wire delivery system. The stent is cut from a nitinol tube into an open lattice design and is designed with tantalum radiopaque markers at the proximal and distal ends of the stent. Upon deployment, the stent achieves its predetermined diameter and exerts a constant, gentle outward force to establish patency in the biliary ducts.
- Status
- Terminated
- Initiated
- 2006-02-27
- Posted
- 2006-03-18
- Terminated
- 2006-12-26
- Reason
- Mispackaging-Protege GPS SER6-9-60-125 devices were incorrectly packaged as SER6-7-80-135. Diameter is larger than expected.
- Root cause
- Other
- Action
- An ev3 field representative will notify affected consignees in person beginning 2/27/2006. Product retrieval or use of the device will be documented using on a ''Field Action Form''. Account credit or product from an unaffected lot number will be provided.
- Quantity
- 4
- Distribution
- GA, OK and TX
- Lot, serial or model codes
- Lot 1211723, expiration date Nov 2007
- 510(k) numbers
- ["K011806"]
- PMA numbers
- not on file
- Product code
- FGE
- Firm FEI
- 2183870
- Firm city
- Plymouth
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28