Philips Medical Systems (Cleveland) Inc: Philips Spectral CT Applications supports viewing and analysis of images at energies selected from the available spectrum in order to provide information about the chemical composition of the body materials and/or contrast agents.
- Recall number
- Z-0706-2017
- Recalling firm
- Philips Medical Systems (Cleveland) Inc
- Product
- Philips Spectral CT Applications supports viewing and analysis of images at energies selected from the available spectrum in order to provide information about the chemical composition of the body materials and/or contrast agents.
- Status
- Terminated
- Initiated
- 2016-11-14
- Posted
- not on file
- Terminated
- 2018-07-02
- Reason
- A software defect is causing issues with the IQon Spectral CT and Philips Spectral CT Applications while filming, saving and annotating images
- Root cause
- Software design
- Action
- A Field Safety Notice (Attachment #1 Field Safety Notice FSN 16-061-72800659 and FSN 88100051), will be sent to US consignees via certified mail. Philips Key Market representatives will distribute the notice internationally, based on local regulations. Additionally, a Field Service Engineer will contact each site to schedule a time to implement the Field Change Order.
- Quantity
- 20
- Distribution
- Software was distributed in the following states: AR, IN, LA, MA, MN, OR and TN. Softward was also shipped to the following countries: Belgium, Denmark, France, Germany, Japan, Korea, Switzerland and United Kingdom.
- Lot, serial or model codes
- 50001 50002 50004 50005 50006 50008 50009 50100 50101 50102 50103 430000 430001 430002 430003 430004 430006 430007 430501 430502
- 510(k) numbers
- ["K150665"]
- PMA numbers
- not on file
- Product code
- JAK
- Firm FEI
- 1525965
- Firm city
- Cleveland
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28