FDA Medical Device Recalls (openFDA)
Diasorin Inc.: Version 7 Software for wash program on Bio Tek Automated Microplate Washer Model ELx50 which performs the washing steps for various programmed mictotiter plate assays.
- Recall number
- Z-0703-05
- Recalling firm
- Diasorin Inc.
- Product
- Version 7 Software for wash program on Bio Tek Automated Microplate Washer Model ELx50 which performs the washing steps for various programmed mictotiter plate assays.
- Status
- Terminated
- Initiated
- 2004-05-20
- Posted
- 2005-04-13
- Terminated
- 2006-07-22
- Reason
- When software was updated for the Washer program of the Bio Tek Automated Microplate Washer Model ELx50, the software did not automatically clear old versions of the wash program when new ones were added.
- Root cause
- Other
- Action
- Customers were notified by phone and informed of the problem and recommended users to use only the newer wash program which is identified is a 'W' in front of the program.
- Quantity
- 9
- Distribution
- CA, FL, NH, PA, SD, TX, WA
- Lot, serial or model codes
- version 7
- 510(k) numbers
- ["K884125"]
- PMA numbers
- not on file
- Product code
- JTC
- Firm FEI
- 2182595
- Firm city
- Stillwater
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28