Boston Scientific Corporation: IntellaNav MiFi Open-Irrigated Ablation Catheter Product Usage: Open Irrigated ablation catheters, when used with a compatible Radiofrequency (RF) Controller and Irrigation Pump, are indicated for: Cardiac electrophysiological mapping Delivering diagnostic pacing stimuli RF ablation of sustained or recurrent type 1 Atrial Flutter in patients age 18 years or older Treatment of drug refractory, recurrent, symptomatic, paroxysmal atrial fibrillation (PAF) in patients age 18 years or older, when used with a compatible mapping system

Recall number
Z-0701-2019
Recalling firm
Boston Scientific Corporation
Product
IntellaNav MiFi Open-Irrigated Ablation Catheter Product Usage: Open Irrigated ablation catheters, when used with a compatible Radiofrequency (RF) Controller and Irrigation Pump, are indicated for: Cardiac electrophysiological mapping Delivering diagnostic pacing stimuli RF ablation of sustained or recurrent type 1 Atrial Flutter in patients age 18 years or older Treatment of drug refractory, recurrent, symptomatic, paroxysmal atrial fibrillation (PAF) in patients age 18 years or older, when used with a compatible mapping system
Status
Open, Classified
Initiated
2018-11-28
Posted
not on file
Terminated
not on file
Reason
Supplier had changed its resin sub-suppliers three (3) times without notifying BSC. These changes resulted in slight variations in resin material compositions which were not reflected in BSC s internal product documentation. Not approved in customer document.
Root cause
Nonconforming Material/Component
Action
On November 28, 2018, Boston Scientific sent Urgent Medical Device Removal - Immediate Action Required letters to their Japanese consignees requesting them to check and segregate any of the recalled product. The letter also asked the consignees to conduct sub-recalls, if necessary, and to return The Reply Verification Tracking Form.
Quantity
1,282 (all three types in Japan)
Distribution
Japan
Lot, serial or model codes
IntellaNav Open-Irrigated Ablation Catheter Material Number: M004R96200 Batch: 22208397 Expiration Date: 6/4/2021 Batch: 22389516 Expiration Date: 7/16/2021 Batch: 22525798 Expiration Date: 2/14/2020
510(k) numbers
not on file
PMA numbers
["P150005S008"]
Product code
OAD
Firm FEI
3004859241
Firm city
Marlborough
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register