Skip to the content

FDA Medical Device Recalls (openFDA)

Neoprobe Corp: Product is CSN00200 4MHz blood flowmeter probe used with the Quantix/OR system.

Recall number
Z-0701-05
Recalling firm
Neoprobe Corp
Product
Product is CSN00200 4MHz blood flowmeter probe used with the Quantix/OR system.
Status
Terminated
Initiated
2004-12-13
Posted
2005-04-13
Terminated
2007-07-21
Reason
The probe's faceplate seal can fail and allow penetration of cleaning fluids during the cleaning and soaking cycle between uses or ingress of biohazard fluids during surgical procedures.
Root cause
Other
Action
The recalling firm notified cutomers by telepone and e mail beginning on 12/2/04.
Quantity
19 devices
Distribution
Domestic accounts: Boston University Medical Center, 88 E. Newton St., B-402, Boston, MA 02118; Criticor, 2943 Pullman St., Santa Ana, CA 92705; Ilex Medical, 370 Cypress Creek Circle, Olsmar, FL 34677; Peerless Medical Corp., 705 Summer Oak Dr., Ellisville, MO 63021; Rolomac International, 1836 NE 213 Ln., Miami FL, 33179; University of Miami, PO Box 01690 (R-114), Miami, FL 33101. Foreign accounts: Asia Health Care Co Ltc., 4 FL-3, No 601, Chung Cheng Rd., Taipei, Taiwan, R.O.C.; Century Medical Inc., 1-6-4 Osaki, Shinagawa-KU, Tokyo, Japan 141-8588; Farmplex Cardic, Sheldare Enterprises, Ltd., 20-22 Bedford Row, London, United Kingdom; Integmed Ltd., 51 Hung To Road Kwun Tong, Kowloon, Hong Kong; Reysas Medical Corp., Hulya Sokal No. 21 GOP, 06700 Ankara, Turkey; Rahn AG, Dorflistrasse 12, CH-8050, Zurici, Switzerland.
Lot, serial or model codes
All serial numbers.
510(k) numbers
["K030357"]
PMA numbers
not on file
Product code
DPW
Firm FEI
3009548429
Firm city
Dublin
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register