Skip to the content

FDA Medical Device Recalls (openFDA)

Icotec Ag: VADER pedicle system, torque wrench, Catalog Number 42-702

Recall number
Z-0700-2024
Recalling firm
Icotec Ag
Product
VADER pedicle system, torque wrench, Catalog Number 42-702
Status
Open, Classified
Initiated
2023-11-28
Posted
2024-01-10
Terminated
not on file
Reason
icotec has received notification of one revision that had to be carried out because nut screws had become loose. During the investigation of the device involved, it was detected that the torque limiting function of the wrench used to tighten the nut screws during implantation was not functioning correctly. Due to the malfunction, the nut screws may not have been tightened sufficiently.
Root cause
Other
Action
icotec ag issued an URGENT Medical Device Recall notice to its consignees beginning on 11/28/2023 via email. The notice explained the problem with the device, the risk to health, and informed the medical facility that the affected product had been removed prior to the notice. Decision on whether patient-level follow-up is required should be made by the physician. Replacements will be provided where applicable. The firm expanded this recall on 12/27/2023 (notices delivered by US mail) to include additional lots and one additional product. Contact for questions: Laura Husken, laura.huesken@icotec.ch, Tel.: +41 78 438 78 73
Quantity
79 devices (US only)
Distribution
US: AZ. MI, OH, CA, CO, TX, NY, NC, DE, PA
Lot, serial or model codes
19/01-001, 19/01-013, 19/01-014, 19/01-015, 190830-10, 190830-23, 190830-42, 190830-44, 190830-45, 190830-50, 190830-51, 190830-52, 200131-54, 200131-56, 200131-62, 200131-64, 200131-68, 200131-69, 200131-70, 200131-72, 200131-73, 200131-74, 200221-040R1, 200424-01, 200424-03, 200424-04, 200424-05, 200424-06, 200424-10, 200424-11, 200424-12, 200424-13, 200424-14, 200424-15, 200424-16, 200424-17, 200424-20, 200424-21, 200424-22, 200424-24, 200424-25, 200424-27, 200424-29, 200424-32, 200424-34, 200424-35, 200605-051R1, 200605-052R1, 201002-073R1, 210915-005, 210915-009, 210915-010, 210915-011, 210915-012, 220610-001, 220610-003, 220610-008, 220610-019, 220610-020, 220610-021, 220610-025, 220610-026, 220610-030, 230113-003, 230113-005, 230113-007, 230113-008, 230113-009, 230113-010, 230113-011, 230113-012, 230113-013, 230113-014, 230224-029, 230224-034, 230426-038, 230426-039, 230426-042, 230426-043
510(k) numbers
["K222789"]
PMA numbers
not on file
Product code
PML
Firm FEI
3006493760
Firm city
Altstatten Sg
Firm state
not on file
Firm country
Switzerland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register