Boston Scientific Corporation: IntellaTip MiFi Open-Irrigated Ablation Catheter Product Usage: Open Irrigated ablation catheters, when used with a compatible Radiofrequency (RF) Controller and Irrigation Pump, are indicated for: Cardiac electrophysiological mapping Delivering diagnostic pacing stimuli RF ablation of sustained or recurrent type 1 Atrial Flutter in patients age 18 years or older Treatment of drug refractory, recurrent, symptomatic, paroxysmal atrial fibrillation (PAF) in patients age 18 years or older, when used with a compatible mapping system

Recall number
Z-0700-2019
Recalling firm
Boston Scientific Corporation
Product
IntellaTip MiFi Open-Irrigated Ablation Catheter Product Usage: Open Irrigated ablation catheters, when used with a compatible Radiofrequency (RF) Controller and Irrigation Pump, are indicated for: Cardiac electrophysiological mapping Delivering diagnostic pacing stimuli RF ablation of sustained or recurrent type 1 Atrial Flutter in patients age 18 years or older Treatment of drug refractory, recurrent, symptomatic, paroxysmal atrial fibrillation (PAF) in patients age 18 years or older, when used with a compatible mapping system
Status
Open, Classified
Initiated
2018-11-28
Posted
not on file
Terminated
not on file
Reason
Supplier had changed its resin sub-suppliers three (3) times without notifying BSC. These changes resulted in slight variations in resin material compositions which were not reflected in BSC s internal product documentation. Not approved in customer document.
Root cause
Nonconforming Material/Component
Action
On November 28, 2018, Boston Scientific sent Urgent Medical Device Removal - Immediate Action Required letters to their Japanese consignees requesting them to check and segregate any of the recalled product. The letter also asked the consignees to conduct sub-recalls, if necessary, and to return The Reply Verification Tracking Form.
Quantity
1,282 (all three types in Japan)
Distribution
Japan
Lot, serial or model codes
IntellaTip MiFi Open-Irrigated Ablation Catheter Material Number: M004PM96200 Batch: 20368736 Expiration Date: 3/5/2020 Batch: 20531648 Expiration Date: 4/16/2020 Batch: 20560477 Expiration Date: 4/23/2020 Batch: 20566637 Expiration Date: 4/24/2020 Batch: 20628704 Expiration Date: 5/9/2020 Batch: 20628705 Expiration Date: 5/10/2020 Batch: 20763900 Expiration Date: 6/12/2020 Batch: 20769113 Expiration Date: 6/13/2020 Batch: 21318000 Expiration Date: 10/29/2020 Batch: 21318001 Expiration Date: 10/29/2020 Batch: 21357647 Expiration Date: 11/10/2020 Batch: 21463186 Expiration Date: 12/3/2020 Batch: 21543193 Expiration Date: 12/20/2020 Batch: 21574774 Expiration Date: 1/4/2021 Batch: 22195929 Expiration Date: 5/31/2021 Batch: 22202991 Expiration Date: 6/3/2021 Batch: 22290601 Expiration Date: 6/3/2021 Batch: 22305067 Expiration Date: 6/25/2021 Batch: 22315782 Expiration Date: 6/27/2021 Batch: 22317356 Expiration Date: 6/28/2021 Batch: 22327178 Expiration Date: 7/1/2021 Batch: 22329367 Expiration Date: 7/2/2021 Batch: 22330626 Expiration Date: 7/2/2021 Batch: 22371866 Expiration Date: 7/12/2021 Batch: 22371868 Expiration Date: 7/12/2021 Batch: 22379905 Expiration Date: 7/15/2021 Batch: 22435929 Expiration Date: 7/26/2021 Batch: 22435930 Expiration Date: 7/26/2021 Batch: 22438118 Expiration Date: 7/27/2021 Batch: 22455198 Expiration Date: 7/31/2021 Batch: 22499957 Expiration Date: 8/10/2021 Material Number: M004PM9620K20 Batch: 20371559 Expiration Date: 3/6/2020 Batch: 20542305 Expiration Date: 4/18/2020 Batch: 20548241 Expiration Date: 4/19/2020 Batch: 20639975 Expiration Date: 5/14/2020 Batch: 20677314 Expiration Date: 5/21/2020 Batch: 20772834 Expiration Date: 6/14/2020 Batch: 22321772 Expiration Date: 6/29/2021 Batch: 22326857 Expiration Date: 7/1/2021 Material Number: M004PM9620N40 Batch: 20371742 Expiration Date: 3/8/2020 Batch: 20463744 Expiration Date: 3/28/2020 Batch: 20622763 Expiration Date: 5/8/2020 Batch: 20775597 Expiration Date: 6/14/2020 Batch: 22340824 Expiration Date: 7/4/2021 Batch: 22371871 Expiration Date: 7/12/2021 Batch: 22374909 Expiration Date: 7/13/2021
510(k) numbers
not on file
PMA numbers
["P150005S005"]
Product code
OAD
Firm FEI
3004859241
Firm city
Marlborough
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register