FDA Medical Device Recalls (openFDA)
Boston Scientific Corporation: IntellaTip MiFi Open-Irrigated Ablation Catheter Product Usage: Open Irrigated ablation catheters, when used with a compatible Radiofrequency (RF) Controller and Irrigation Pump, are indicated for: Cardiac electrophysiological mapping Delivering diagnostic pacing stimuli RF ablation of sustained or recurrent type 1 Atrial Flutter in patients age 18 years or older Treatment of drug refractory, recurrent, symptomatic, paroxysmal atrial fibrillation (PAF) in patients age 18 years or older, when used with a compatible mapping system
- Recall number
- Z-0700-2019
- Recalling firm
- Boston Scientific Corporation
- Product
- IntellaTip MiFi Open-Irrigated Ablation Catheter Product Usage: Open Irrigated ablation catheters, when used with a compatible Radiofrequency (RF) Controller and Irrigation Pump, are indicated for: Cardiac electrophysiological mapping Delivering diagnostic pacing stimuli RF ablation of sustained or recurrent type 1 Atrial Flutter in patients age 18 years or older Treatment of drug refractory, recurrent, symptomatic, paroxysmal atrial fibrillation (PAF) in patients age 18 years or older, when used with a compatible mapping system
- Status
- Open, Classified
- Initiated
- 2018-11-28
- Posted
- not on file
- Terminated
- not on file
- Reason
- Supplier had changed its resin sub-suppliers three (3) times without notifying BSC. These changes resulted in slight variations in resin material compositions which were not reflected in BSC s internal product documentation. Not approved in customer document.
- Root cause
- Nonconforming Material/Component
- Action
- On November 28, 2018, Boston Scientific sent Urgent Medical Device Removal - Immediate Action Required letters to their Japanese consignees requesting them to check and segregate any of the recalled product. The letter also asked the consignees to conduct sub-recalls, if necessary, and to return The Reply Verification Tracking Form.
- Quantity
- 1,282 (all three types in Japan)
- Distribution
- Japan
- Lot, serial or model codes
- IntellaTip MiFi Open-Irrigated Ablation Catheter Material Number: M004PM96200 Batch: 20368736 Expiration Date: 3/5/2020 Batch: 20531648 Expiration Date: 4/16/2020 Batch: 20560477 Expiration Date: 4/23/2020 Batch: 20566637 Expiration Date: 4/24/2020 Batch: 20628704 Expiration Date: 5/9/2020 Batch: 20628705 Expiration Date: 5/10/2020 Batch: 20763900 Expiration Date: 6/12/2020 Batch: 20769113 Expiration Date: 6/13/2020 Batch: 21318000 Expiration Date: 10/29/2020 Batch: 21318001 Expiration Date: 10/29/2020 Batch: 21357647 Expiration Date: 11/10/2020 Batch: 21463186 Expiration Date: 12/3/2020 Batch: 21543193 Expiration Date: 12/20/2020 Batch: 21574774 Expiration Date: 1/4/2021 Batch: 22195929 Expiration Date: 5/31/2021 Batch: 22202991 Expiration Date: 6/3/2021 Batch: 22290601 Expiration Date: 6/3/2021 Batch: 22305067 Expiration Date: 6/25/2021 Batch: 22315782 Expiration Date: 6/27/2021 Batch: 22317356 Expiration Date: 6/28/2021 Batch: 22327178 Expiration Date: 7/1/2021 Batch: 22329367 Expiration Date: 7/2/2021 Batch: 22330626 Expiration Date: 7/2/2021 Batch: 22371866 Expiration Date: 7/12/2021 Batch: 22371868 Expiration Date: 7/12/2021 Batch: 22379905 Expiration Date: 7/15/2021 Batch: 22435929 Expiration Date: 7/26/2021 Batch: 22435930 Expiration Date: 7/26/2021 Batch: 22438118 Expiration Date: 7/27/2021 Batch: 22455198 Expiration Date: 7/31/2021 Batch: 22499957 Expiration Date: 8/10/2021 Material Number: M004PM9620K20 Batch: 20371559 Expiration Date: 3/6/2020 Batch: 20542305 Expiration Date: 4/18/2020 Batch: 20548241 Expiration Date: 4/19/2020 Batch: 20639975 Expiration Date: 5/14/2020 Batch: 20677314 Expiration Date: 5/21/2020 Batch: 20772834 Expiration Date: 6/14/2020 Batch: 22321772 Expiration Date: 6/29/2021 Batch: 22326857 Expiration Date: 7/1/2021 Material Number: M004PM9620N40 Batch: 20371742 Expiration Date: 3/8/2020 Batch: 20463744 Expiration Date: 3/28/2020 Batch: 20622763 Expiration Date: 5/8/2020 Batch: 20775597 Expiration Date: 6/14/2020 Batch: 22340824 Expiration Date: 7/4/2021 Batch: 22371871 Expiration Date: 7/12/2021 Batch: 22374909 Expiration Date: 7/13/2021
- 510(k) numbers
- not on file
- PMA numbers
- ["P150005S005"]
- Product code
- OAD
- Firm FEI
- 3004859241
- Firm city
- Marlborough
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as