Skip to the content

FDA Medical Device Recalls (openFDA)

Ion Beam Applications S.A.: Proteus 235 Product Usage: The Proton Therapy System - Proteus 235 (brand names: Proteus Plus and Proteus ONE) is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. The PTS may include a fixed small beam treatment room dedicated to the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation localized to the head and neck.

Recall number
Z-0698-2019
Recalling firm
Ion Beam Applications S.A.
Product
Proteus 235 Product Usage: The Proton Therapy System - Proteus 235 (brand names: Proteus Plus and Proteus ONE) is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. The PTS may include a fixed small beam treatment room dedicated to the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation localized to the head and neck.
Status
Terminated
Initiated
2018-12-12
Posted
not on file
Terminated
2020-04-15
Reason
There is a risk of detachment of the Compact Gantry Rolling Floor.
Root cause
Device Design
Action
On December 12, 2018, the firm notified its customers of the recall via an Urgent Medical Device Correction letter. Customers were informed of the product issue. Customers were reminded that nobody should stand on the rolling floor when the gantry is rotating, as specified in the Clinical User Guide. The recalling firm stated they would take the following actions: Short-term action: IBA will replace the rod end bearings attaching the cGRF to the nozzle via a preventive maintenance on all ProteusONE sites. Long-term action: A new design of the anchoring system of the cGRF to the nozzle will be deployed on all impacted sites via an upgrade kit by end of October 2019. Customers with questions should contact Sylviane BERGER Vigilance@iba-group.com +32 10 203 787, or Helpdesk +32 2 507 20 81 (available 24/7)
Quantity
7
Distribution
Worldwide Distribution - US Nationwide in the states of LA and MI. Foreign distribution to France, Japan, and UK.
Lot, serial or model codes
SAT.123, SBF.101, SBF.103, SBF.104, SBF.105, SBF.107, SBF.109
510(k) numbers
["K091629"]
PMA numbers
not on file
Product code
LHN
Firm FEI
3000256071
Firm city
Louvain-la-neuve
Firm state
not on file
Firm country
Belgium

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register