Skip to the content

FDA Medical Device Recalls (openFDA)

Microbiologics Inc: Microbiologics QC Sets and Panels: Helix Elite. Extended Enteric Bacterial Verification Panel (Inactivated Pellet) includes 21 pellets each comprised of the following 8 organisms: 1. S. Typhimurium derived from NCTC 74, 2. S. sonnei derived from NCTC 12984, 3. E. coli derived from CDC 2010C-3114 (STEC), 4. C. jejuni derived from NCTC 11322, 5. P. shigelloides derived from NCTC 10363, 6. Y. enterocolitica derived from NCTC 11174, 7. E. coli derived from NCTC 11602 (ETEC), 8. V. parahaemolyticus derived from NCTC 13209

Recall number
Z-0697-2022
Recalling firm
Microbiologics Inc
Product
Microbiologics QC Sets and Panels: Helix Elite. Extended Enteric Bacterial Verification Panel (Inactivated Pellet) includes 21 pellets each comprised of the following 8 organisms: 1. S. Typhimurium derived from NCTC 74, 2. S. sonnei derived from NCTC 12984, 3. E. coli derived from CDC 2010C-3114 (STEC), 4. C. jejuni derived from NCTC 11322, 5. P. shigelloides derived from NCTC 10363, 6. Y. enterocolitica derived from NCTC 11174, 7. E. coli derived from NCTC 11602 (ETEC), 8. V. parahaemolyticus derived from NCTC 13209
Status
Terminated
Initiated
2017-06-06
Posted
not on file
Terminated
2023-01-17
Reason
Real-time shelf life testing failed at 24 months
Root cause
Process design
Action
Microbiologics notified customers on about 06/06/2017 via email containing a letter. The letters indicated the affected products and instructed customers to update their records with the corrected expiration date, which was provided for each affected lot.
Quantity
8 units
Distribution
Worldwide distribution - US Nationwide distribution in the states of AK, CA, DC, FL, IA, ID, KS, LA, MD, MN, MO, NE, NH, NJ, NY, OH, OK, PA, SC, TX, VA, WA, WV and the countries of Australia, Belgium, Canada, India, United Kingdom.
Lot, serial or model codes
Catalog Number: 8191, Lot Number: 8191-01, UDI: 70845357041240
510(k) numbers
not on file
PMA numbers
not on file
Product code
OHQ
Firm FEI
2150138
Firm city
Saint Cloud
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register