FDA Medical Device Recalls (openFDA)
Phadia Ab: EliA SmDP Well, Article Number 14-5624-01 Product Usage: EliA SmDP is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Sm in human serum and plasma (EDTA, citrate) as an aid in the clinical diagnosis of systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA SmDP uses the EliA IgG method on the instruments Phadia 100 and Phadia 250.
- Recall number
- Z-0697-2016
- Recalling firm
- Phadia Ab
- Product
- EliA SmDP Well, Article Number 14-5624-01 Product Usage: EliA SmDP is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Sm in human serum and plasma (EDTA, citrate) as an aid in the clinical diagnosis of systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA SmDP uses the EliA IgG method on the instruments Phadia 100 and Phadia 250.
- Status
- Terminated
- Initiated
- 2015-12-28
- Posted
- 2016-01-28
- Terminated
- 2016-08-16
- Reason
- All equivocal and positive results (greater or equal 7 EliA U/ml) on EliA SmDP well lot 0018 may be incorrect and must be considered invalid. Patient samples in complaint investigations caused unspecific signals up to 22 U/ml on EliA SmDP well lot 0018. The nonspecific signals are not caused by anti-Sm antibodies, nor by streptavidin antibodies.
- Root cause
- Nonconforming Material/Component
- Action
- On 12/28/2015, URGENT MEDICAL DEVICE RECALL notifications were sent to the affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions call 1-800-3464364, option #2.
- Quantity
- 76
- Distribution
- US Nationwide in the states of NC, GA, NJ, NY, PA, TX, IL, MI, MT, NM, and OR.
- Lot, serial or model codes
- Product number: 14-5624-01 Lot: 0018 Exp date: 11/30/2016
- 510(k) numbers
- ["K132631"]
- PMA numbers
- not on file
- Product code
- LKP
- Firm FEI
- 3002807876
- Firm city
- Uppsala
- Firm state
- not on file
- Firm country
- Sweden
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28