Skip to the content

FDA Medical Device Recalls (openFDA)

Phadia Ab: EliA SmDP Well, Article Number 14-5624-01 Product Usage: EliA SmDP is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Sm in human serum and plasma (EDTA, citrate) as an aid in the clinical diagnosis of systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA SmDP uses the EliA IgG method on the instruments Phadia 100 and Phadia 250.

Recall number
Z-0697-2016
Recalling firm
Phadia Ab
Product
EliA SmDP Well, Article Number 14-5624-01 Product Usage: EliA SmDP is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Sm in human serum and plasma (EDTA, citrate) as an aid in the clinical diagnosis of systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA SmDP uses the EliA IgG method on the instruments Phadia 100 and Phadia 250.
Status
Terminated
Initiated
2015-12-28
Posted
2016-01-28
Terminated
2016-08-16
Reason
All equivocal and positive results (greater or equal 7 EliA U/ml) on EliA SmDP well lot 0018 may be incorrect and must be considered invalid. Patient samples in complaint investigations caused unspecific signals up to 22 U/ml on EliA SmDP well lot 0018. The nonspecific signals are not caused by anti-Sm antibodies, nor by streptavidin antibodies.
Root cause
Nonconforming Material/Component
Action
On 12/28/2015, URGENT MEDICAL DEVICE RECALL notifications were sent to the affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions call 1-800-3464364, option #2.
Quantity
76
Distribution
US Nationwide in the states of NC, GA, NJ, NY, PA, TX, IL, MI, MT, NM, and OR.
Lot, serial or model codes
Product number: 14-5624-01 Lot: 0018 Exp date: 11/30/2016
510(k) numbers
["K132631"]
PMA numbers
not on file
Product code
LKP
Firm FEI
3002807876
Firm city
Uppsala
Firm state
not on file
Firm country
Sweden

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register