Skip to the content

FDA Medical Device Recalls (openFDA)

DePuy Mitek, Inc., a Johnson & Johnson Co.: 1113 Gravity Tube Set Pk/24 -SH Product Usage: The FMS Gravity Tube Set is intended to deliver irrigation fluid from irrigation containers to the operative site during arthroscopic procedures.

Recall number
Z-0695-2018
Recalling firm
DePuy Mitek, Inc., a Johnson & Johnson Co.
Product
1113 Gravity Tube Set Pk/24 -SH Product Usage: The FMS Gravity Tube Set is intended to deliver irrigation fluid from irrigation containers to the operative site during arthroscopic procedures.
Status
Terminated
Initiated
2017-08-31
Posted
not on file
Terminated
2019-04-09
Reason
Leakage of FMS gravity tub sets (281113) at the bulb and tube connection
Root cause
Nonconforming Material/Component
Action
On September 6, 2017 an URGENT MEDICAL DEVICE RECALL letter was issued to customers titled "FMS Gravity Tubing" requesting that customers cease further distribution or use, quarantine all remaining products and contact sales consultant for the coordination of return and replacement. Questions or concerns can be directed to the recall coordinator at 508-828-6609
Quantity
15,678
Distribution
US Nationwide Distribution
Lot, serial or model codes
Lot codes: 138, 194, 633, 1067, 1741, 2576, 2608, and 2645
510(k) numbers
not on file
PMA numbers
not on file
Product code
JCY
Firm FEI
1221934
Firm city
Raynham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register