Skip to the content

FDA Medical Device Recalls (openFDA)

Gambro Renal Products, Inc.: Gambro Prismaflex Continuous Renal Replacement System, Catalog Number 602314700, Gambro Lundia AB, Sweden.

Recall number
Z-0695-2007
Recalling firm
Gambro Renal Products, Inc.
Product
Gambro Prismaflex Continuous Renal Replacement System, Catalog Number 602314700, Gambro Lundia AB, Sweden.
Status
Terminated
Initiated
2007-01-04
Posted
2007-03-31
Terminated
2012-01-03
Reason
Under certain conditions, * an excessive amount of anticoagulant may be infused into the patient. (* interruption of a self-test by an alarm and obstruction of the access lines)
Root cause
Device Design
Action
Customers were notified by letter on 12/19/2006. They were told that Gambro would schedule the required upgrades to their equipment and not to use the current syringe for delivering anticoagulant.
Quantity
153 units (Domestic)
Distribution
Nationwide distribution -- including states of AL, CA, FL, GA, IN, MD, ME, MI, MS, NC, NH, NY, OH, PA, SC, TX, VA, and WI.
Lot, serial or model codes
Serial numbers: PA0101 thru PA0681, PA1001 thru PA1774.
510(k) numbers
["K041005"]
PMA numbers
not on file
Product code
KDI
Firm FEI
1713683
Firm city
Lakewood
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register