FDA Medical Device Recalls (openFDA)
Perkinelmer: Microplate, truncated V-bottomed lots included in the above listed products Product Number: 1380 5465. DBS Microplate is a general purpose 96 well microplate used in the preparation and examination of specimens from the human body.
- Recall number
- Z-0690-2013
- Recalling firm
- Perkinelmer
- Product
- Microplate, truncated V-bottomed lots included in the above listed products Product Number: 1380 5465. DBS Microplate is a general purpose 96 well microplate used in the preparation and examination of specimens from the human body.
- Status
- Terminated
- Initiated
- 2012-11-26
- Posted
- 2013-01-16
- Terminated
- 2014-03-07
- Reason
- Truncated V-Bottomed Microplate Wells used in assays contain defective wells.
- Root cause
- Process control
- Action
- Perkin Elmer initiated the recall on November 23, 2012. The Recall Letter will be delivered to the end customer by email facilitated through our PerkinElmer Technical Support group located in Akron Ohio. Technical Support will be asked to make sure the customers receive the Recall Letter and implement the recommended precautions: - Inspect the wells of row A and row H of the affected microplate lots for possible holes and cracks prior to use. - If a hole or crack is visually detected in a well of the microplate, the well should not be used for testing. - If a hole or crack is not visually detected in a well of the plate, it is required during use of the well, the volume of solution in the well should be visually confirmed to be consistent before and after incubation. Please contact PerkinEimer Technical Support for further information (800) 321 -9632, select prompt 2.
- Quantity
- not on file
- Distribution
- Nationwide Distribution.
- Lot, serial or model codes
- Lot Numbers: 616792 , 617141 , 617980 , 618319, 618664, 618831, 619175
- 510(k) numbers
- ["K071649"]
- PMA numbers
- not on file
- Product code
- not on file
- Firm FEI
- 3006159931
- Firm city
- Waltham
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28