Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic CardioVascular: Medtronic brand AneuRx AAAdvantage Endovascular Stent Graft with the Xcelerant Delivery System; Model Numbers Affected: All AAAdvantage models including: AEXC202040, IEXC121255, ILXC1212115, ILXC151585, AEXC222240, IEXC131355, ILXC1212135, ILXC1616115, AEXC242440, IEXC141455, ILXC121285, ILXC1616135, AEXC262640, IEXC151555, ILXC1313115, ILXC161685, AEXC282840, IEXC161655, ILXC1313135, ILXC1620115, BFXC2012135, IEXC162085, ILXC131385, ILXC1620135, BFXC2012165, IEXC181855, ILXC1414115, ILXC1818115, BFXC2213135, IEXC182285, ILXC1414135, ILXC1818135, BFXC2213165, IEXC182485, ILXC141485, ILXC181885, BFXC2414135, IEXC202055, ILXC1515115, ILXC1822115, BFXC2414165, ILXC1515135, ILXC1822135, BFXC2615135, ILXC1824115, BFXC2615165, ILXC1824135, BFXC2816135, ILXC2020115, BFXC2816165, ILXC2020135, ILXC202085; Product is manufactured and distributed by Medtronic, Inc., Santa Rosa, CA 95403

Recall number
Z-0689-2008
Recalling firm
Medtronic CardioVascular
Product
Medtronic brand AneuRx AAAdvantage Endovascular Stent Graft with the Xcelerant Delivery System; Model Numbers Affected: All AAAdvantage models including: AEXC202040, IEXC121255, ILXC1212115, ILXC151585, AEXC222240, IEXC131355, ILXC1212135, ILXC1616115, AEXC242440, IEXC141455, ILXC121285, ILXC1616135, AEXC262640, IEXC151555, ILXC1313115, ILXC161685, AEXC282840, IEXC161655, ILXC1313135, ILXC1620115, BFXC2012135, IEXC162085, ILXC131385, ILXC1620135, BFXC2012165, IEXC181855, ILXC1414115, ILXC1818115, BFXC2213135, IEXC182285, ILXC1414135, ILXC1818135, BFXC2213165, IEXC182485, ILXC141485, ILXC181885, BFXC2414135, IEXC202055, ILXC1515115, ILXC1822115, BFXC2414165, ILXC1515135, ILXC1822135, BFXC2615135, ILXC1824115, BFXC2615165, ILXC1824135, BFXC2816135, ILXC2020115, BFXC2816165, ILXC2020135, ILXC202085; Product is manufactured and distributed by Medtronic, Inc., Santa Rosa, CA 95403
Status
Terminated
Initiated
2007-12-10
Posted
2008-04-08
Terminated
2008-07-28
Reason
Sterility compromised: Device sterility may be compromised as evidenced by a loss of outer package integrity.
Root cause
Package design/selection
Action
Medtronic sent Urgent Communication - Medical Device Correction Advisory Notice letters to all its direct consignees on 12/7/07, via Federal Express, informing them of the affected product and providing instructions on avoiding contamination of the sterile surgical field with these devices.
Quantity
7,000
Distribution
Nationwide Distribution.
Lot, serial or model codes
All codes
510(k) numbers
not on file
PMA numbers
["P990020"]
Product code
MIH
Firm FEI
3001452571
Firm city
Santa Rosa
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register