FDA Medical Device Recalls (openFDA)
OEC Medical Systems, Inc: ENTrak Navigation and Visualization System
- Recall number
- Z-0687-05
- Recalling firm
- OEC Medical Systems, Inc
- Product
- ENTrak Navigation and Visualization System
- Status
- Terminated
- Initiated
- 2004-12-13
- Posted
- 2005-04-08
- Terminated
- 2011-07-08
- Reason
- Enhanced sterility testing revealed the sterilization efficacy of accessories ('Sensors') used with certain Navigation and Visualization systems may be compromised.
- Root cause
- Other
- Action
- Customers were notified by letter on 12/13/04.
- Quantity
- 99 units
- Distribution
- Domestic distribution: Nationwide, including 4 military facilities and 4 VA facilities. Foreign distribution: Canada.
- Lot, serial or model codes
- The particular accessories (''Sensors'') involved are: Short Range Transmitter (P/N 1004587 or 1002008); Snap Receiver Cable (P/N 1004069 or 1001989); Extended Range Transmitter (P/N 1004070 or 1003991); and T-Handle Spine Tool Receiver (P/N 1004072).
- 510(k) numbers
- ["K960330"]
- PMA numbers
- not on file
- Product code
- LLZ
- Firm FEI
- 1720753
- Firm city
- Salt Lake City
- Firm state
- UT
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as