MicroPort Orthopedics Inc.: Procotyl (Rim- Lock) Liners: REF PHA04702, REF PHA04704, REF PHA04706, REF PHA04708, REF PHA04710, REF PHA04712, REF PHA04714 For use with the conserve metal head

Recall number
Z-0685-2017
Recalling firm
MicroPort Orthopedics Inc.
Product
Procotyl (Rim- Lock) Liners: REF PHA04702, REF PHA04704, REF PHA04706, REF PHA04708, REF PHA04710, REF PHA04712, REF PHA04714 For use with the conserve metal head
Status
Terminated
Initiated
2016-06-21
Posted
2016-12-01
Terminated
2017-07-01
Reason
Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was found that there was a specific hazard/harm for suspected tissue reaction to metal debris over one percent for modular head CONSERVE(R) systems and DYNASTY(R) metal acetabular liner systems.
Root cause
Device Design
Action
MicroPort sent an Urgent Field Safety Notice dated June 22, 2016, to all affected customers notifying them of the change in the instructions for use on July 7th, 2016 by email. The firm also notified US hospitals where the implantation surgeries occurred within the last twelve (12) months by letter on July 13th, 2016 delivered by FedEx Priority. Customers were asked to complete and return the enclosed form even if they do not have any affected product to return by fax to 1-901-451-6032 or PostMarket@ortho.microport.com. For questions regarding this recall call 901-867-4771.
Quantity
500 units
Distribution
Worldwide Distribution - US (nationwide) Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Columbia, Cyprus, Czech Republic, Denmark, Finland, France, Georgia, Germany, Greece, Iran, Italy, Jamaica, Japan, Lithuania, Malaysia, Philippines, Poland, Portugal, Russia, Slovenia, South Africa, South Korea, Spain, Switzerland, Taiwan, Netherlands, Turkey, Arab Emirates, United Kingdom, and Vietnam.
Lot, serial or model codes
REF PHA04702 Serial Numbers: PHA04702, 087461467, 089909592, 106368684, 106379196, 106381827, 109949208, 129989153, 0201018636, 0301049040   REF PHA04704 Serial Numbers: 010989155, 027373842, 038461468, 057433172, 077453350, 089924904, 106379197, 106380366, 106381751, 0201018637   REF PHA04706 Serial Numbers: 1414640, 010990534, 017400915, 037415861, 039755749, 047420835, 047424536, 077459275, 087461469, 089906466, 0101034747, 106379879, 106381773, 0301084459, 0401100079, 0501122040, 0801189074   REF PHA04708 Serial Numbers: 010990535, 017400568, 037420836, 047424415, 077455537, 087461470, 089901226, 106380367, 106381752, 0311304891, 0401118999, 0601163086, 0711360334   REF PHA04710 Serial Numbers: 1372615, 1413361, 017400673, 037415862, 037420837, 047423789, 047424535, 077457450, 077460608, 089901146, 098695744, 098695746, 106380368, 106381542, 106381753, 108695745, 109944906, 116380977, 119983782, 129990536, 0201016259, 0301076759, 0311303536, 0311305118, 0401089452, 0401121936, 0501131023, 0801187747, 1291000322   REF PHA04712 Serial Numbers: 029695877, 037414678, 047418090, 057426407, 077457328, 098464129, 099906467, 106373846, 106380571, 116389410, 0201058381, 0301084465, 0311303537, 0401117709, 0601163076,  0801188639   REF PHA04714 Serial Numbers: 1414641, 017400569, 029807399, 037415854, 047423790, 077432883, 077445957, 106368690, 0201025145, 0901196483
510(k) numbers
["K100866"]
PMA numbers
not on file
Product code
LZO
Firm FEI
3010536692
Firm city
Arlington
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register