Skip to the content

FDA Medical Device Recalls (openFDA)

E.M.S. - Electro Medical Systems SA: Swiss Lithocast Ultra Ultrasound Handpiece (EMS- PN:EL-236) used in combination with Swiss Lithoclast Ultra System .

Recall number
Z-0684-2007
Recalling firm
E.M.S. - Electro Medical Systems SA
Product
Swiss Lithocast Ultra Ultrasound Handpiece (EMS- PN:EL-236) used in combination with Swiss Lithoclast Ultra System .
Status
Terminated
Initiated
2005-03-16
Posted
2007-03-30
Terminated
2016-02-02
Reason
Handpiece may become dislodged causing cable damage and possible electric shock.
Root cause
Other
Action
EMS Lyon, France (manufacturer) notified the single US Distributor by letter to contact their accounts and return device for replacement. Boston Scientific, Natick, Ma conducted the sub-recall by letter dated 4/22/05 to all USA accounts.
Quantity
160 units
Distribution
Nationwide
Lot, serial or model codes
Units manufactured prior to January 2003
510(k) numbers
["K012445"]
PMA numbers
not on file
Product code
FFK
Firm FEI
3002806910
Firm city
Nyon
Firm state
Vaud
Firm country
Switzerland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register