Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Sensis Vibe Hemo

Recall number
Z-0683-2024
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Sensis Vibe Hemo
Status
Open, Classified
Initiated
2023-11-29
Posted
2024-01-09
Terminated
not on file
Reason
The possibility of the Sensis documentation functionality application to crash.
Root cause
Software design
Action
The recalling firm issued an URGENT: MEDICAL DEVICE CORRECTION notification dated 11/29/23 by mail to consignees to notify them of this recall event. In the case this failure mode occurs during use, consignees are asked to continue the exam by opening a new study for the same patient without adding any Type 1 reporting events. It is instructed that consignees can compensate this by adding free text comments (procedure notes) only and manually inserting the desired information in those comments. The recalling firm will deploy a software update via Update Instruction AX010/23/S to correct the error. A member of the service organization will contact consignees to schedule the update. Consignees can contact the service organization at 1-800-888-7436 to schedule an earlier appointment. The recall notice is to be disseminated to all users of affected products within consignee organizations and others who may need to be informed. Consignees are to return the provided response form by email to recallsandrefusals.team@siemens-healthineers.com within 14 days of receipt.
Quantity
331 units
Distribution
Domestic: CA, CT, & FL.
Lot, serial or model codes
Model No. 11007641; UDI-DI 04056869010199; Serial No. 18009, 106245, 18016, 18014, 18011, 18010, 19001, 106244, 18015.
510(k) numbers
["K223812"]
PMA numbers
not on file
Product code
DQK
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register