FDA Medical Device Recalls (openFDA)
Siemens Medical Solutions USA, Inc: Sensis Vibe Hemo
- Recall number
- Z-0683-2024
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- Sensis Vibe Hemo
- Status
- Open, Classified
- Initiated
- 2023-11-29
- Posted
- 2024-01-09
- Terminated
- not on file
- Reason
- The possibility of the Sensis documentation functionality application to crash.
- Root cause
- Software design
- Action
- The recalling firm issued an URGENT: MEDICAL DEVICE CORRECTION notification dated 11/29/23 by mail to consignees to notify them of this recall event. In the case this failure mode occurs during use, consignees are asked to continue the exam by opening a new study for the same patient without adding any Type 1 reporting events. It is instructed that consignees can compensate this by adding free text comments (procedure notes) only and manually inserting the desired information in those comments. The recalling firm will deploy a software update via Update Instruction AX010/23/S to correct the error. A member of the service organization will contact consignees to schedule the update. Consignees can contact the service organization at 1-800-888-7436 to schedule an earlier appointment. The recall notice is to be disseminated to all users of affected products within consignee organizations and others who may need to be informed. Consignees are to return the provided response form by email to recallsandrefusals.team@siemens-healthineers.com within 14 days of receipt.
- Quantity
- 331 units
- Distribution
- Domestic: CA, CT, & FL.
- Lot, serial or model codes
- Model No. 11007641; UDI-DI 04056869010199; Serial No. 18009, 106245, 18016, 18014, 18011, 18010, 19001, 106244, 18015.
- 510(k) numbers
- ["K223812"]
- PMA numbers
- not on file
- Product code
- DQK
- Firm FEI
- 2240869
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28