Bayer Healthcare, LLC (Diagnostics Division): Brand Name: RapidLab 1200 System, Model 1240, Part Number 05060298; All RapidLab 1200 Systems with Software Versions 1.0, 1.01, or 1.02. --- Common Name: Blood gases, electrolyte and blood pH test system.

Recall number
Z-0682-06
Recalling firm
Bayer Healthcare, LLC (Diagnostics Division)
Product
Brand Name: RapidLab 1200 System, Model 1240, Part Number 05060298; All RapidLab 1200 Systems with Software Versions 1.0, 1.01, or 1.02. --- Common Name: Blood gases, electrolyte and blood pH test system.
Status
Terminated
Initiated
2006-01-20
Posted
2006-03-29
Terminated
2012-05-02
Reason
When stored data is recalled from memory, the results do not agree with the original results. Software anomaly: a calculation error occurs when using the correlation adjustment feature on the RapidLab 1200 Systems with software versions 1.0, 1.01, or 1.02. The correlation feature applies user-defined slope and offset parameters each time stored test results are recalled to the display.
Root cause
Other
Action
Support Bulletins and Customer Bulletins were e-mailed to all Bayer HealthCare LLC Branches on 1/20/06 for distribution to the end user facilities (medical facilities).
Quantity
not on file
Distribution
USA, Australia, Austria, Belgium, Denmark, Finland, France, Germany, Greece, Hong Kong, Italy, Japan, Korea, Malaysia, The Netherlands, Poland, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland.
Lot, serial or model codes
All RapidLab 1200 Systems with Software Versions 1.0, 1.01, or 1.02.
510(k) numbers
["K031560"]
PMA numbers
not on file
Product code
CHL
Firm FEI
2432235
Firm city
Tarrytown
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register