GE Medical Systems, LLC: GE Healthcare, Aestiva MRI, 1006-9310-000. This version of the Datex-Ohmeda 7900 ventilator is used in Datex-Ohmeda Aestiva/5 anesthesia systems. It is a microprocessor based, electronically controlled, pneumatically driven ventilator that provides patient ventilation during surgical procedures. The Aestiva/5 MRI anesthesia system provides the functional feature set offered by the conventional Aestiva/5 to the clinician with the added ability to be used in the MR environment.

Recall number
Z-0679-2016
Recalling firm
GE Medical Systems, LLC
Product
GE Healthcare, Aestiva MRI, 1006-9310-000. This version of the Datex-Ohmeda 7900 ventilator is used in Datex-Ohmeda Aestiva/5 anesthesia systems. It is a microprocessor based, electronically controlled, pneumatically driven ventilator that provides patient ventilation during surgical procedures. The Aestiva/5 MRI anesthesia system provides the functional feature set offered by the conventional Aestiva/5 to the clinician with the added ability to be used in the MR environment.
Status
Terminated
Initiated
2015-12-15
Posted
2016-01-21
Terminated
2016-09-21
Reason
The Drive Gas Check Valve of certain Anesthesia products could become stuck in a fixed open position, which could cause pressure to build in the mechanical ventilation cycle. If this issue is left unresolved, it could result in excessive or prolonged pressure in the patient breathing circuit during ventilation potentially resulting in barotrauma.
Root cause
Nonconforming Material/Component
Action
GE Healthcare sent an Urgent Medical Device Correction letter dated December 15, 2015, to all affected consignees. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. Customers with questions were instructed to contact GE Healthcare Service at 1-800-437-1171 or their local Service Representative.
Quantity
84
Distribution
Worldwide Distribution - US (nationwide including PR except NH, VT, & WY. Internationally to ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BANGLADESH, BELGIUM, BOLIVIA, BOSNIA, BRAZIL, CAMBODIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, ESTONIA, FINLAND, FRANCE, GEORGIA, GERMANY, CHANA, GIBRALTAR, GREECE, GUATEMALA, HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN KENYA, REPUBLIC OF KOREA, KUWAIT, LATVIA, LITHUANIA,MALAYSIA, MALDIVES, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NEW ZEALAND, NICARAGUA, NORWAY, OMAN, PAKISTAN, PERU, PHILIPPINES, POLAND, QATAR, REUNION, ROMANIA, RUSSIAN FEDERATION SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SRI lanka, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, VENEZUELA, VIETNAM.
Lot, serial or model codes
AMTU01048 AMTU01023 AMTU00163 AMTU01050 AMTU00172 AMTU00175 AMTU00176 AMTU01034 AMTU01047 AMTU01055 AMTU01001 AMTU01002 AMTU01021 AMTU01044 AMTU01042 AMTU01015 AMTU00178 AMTU00179 AMTU00162 AMTU01000 AMTU01012 AMTU01041 AMTU00167 AMTU01019 AMTU00177 AMTU01049 AMTU00171 AMTU00164 AMTU01066 AMTU01006 AMTU00168 AMTU01061 AMTU01060 AMTU01056 AMTU01057 AMTU01073 AMTU00154 AMTU00166 AMTU00174 AMTU01005 AMTU01035 AMTU00150 AMTU01024 AMTU01025 AMTU01026 AMTU01027 AMTU01037 AMTU00152 AMTU01040 AMTU01022 AMTU00157 AMTU00158 AMTU00159 AMTU00160 AMTU00170 AMTU01018 AMTU01009 AMTU01032 AMTU01046 AMTU01045 AMTU01020 AMTU01008 AMTU00165 AMTU01013 AMTU01014 AMTU01033 AMTU01016 AMTU01036 AMTU01038 AMTU00151 AMTU00169 AMTU01003 AMTU00161 AMTU01010 AMTU01064 AMTU01011 AMTU01004 AMTU01017 AMTU01031 AMTU01053 AMTU01058 AMTU01059 AMTU01039 AMTU01007
510(k) numbers
["K050055"]
PMA numbers
not on file
Product code
BSZ
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register