The Spineology Group: Fill Tube,straight, Catalog # 300-1001, is used for the delivery of granular autograft or allograft into Spineology''s OptiMesh implant. Fill tube is identified with Spineology, catalog number and lot number printed directly onto the metal.
- Recall number
- Z-0678-05
- Recalling firm
- The Spineology Group
- Product
- Fill Tube,straight, Catalog # 300-1001, is used for the delivery of granular autograft or allograft into Spineology''s OptiMesh implant. Fill tube is identified with Spineology, catalog number and lot number printed directly onto the metal.
- Status
- Terminated
- Initiated
- 2005-01-28
- Posted
- 2005-04-06
- Terminated
- 2005-11-17
- Reason
- Three malfunctions, due to jamming of bone material in the fill tube occurred, which allowed the fill tube to advance more deeply than intended into the surgical site. Fill tubes are used in delivery of bone graft materials.
- Root cause
- Other
- Action
- A Product Update, January 2005 (Vol. 2) was e-mailed to all representatives and also sent by mail on January 28, 2005. Telephone calls to each distributor/representative were made by Spineology Sales & Marketing staff between January 28 and January 31. The Product Update informs customers that they will be receiving new tubes and new platens soon and that any tubes and platens currently in their possession should be returned to Spineology.
- Quantity
- 870
- Distribution
- 11 consignees located in: Arkansas, Arizona, California, Georgia, Illinois, Michigan, Minnesota, North Carolina, New York, Oklahoma and United Kingdom
- Lot, serial or model codes
- UN02001, 71-1057, UN04001, UN04004, and M04041
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LXH
- Firm FEI
- 3003179719
- Firm city
- Stillwater
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28