The Spineology Group: Fill Tube,diverted, Catalog # 300-1002, is used for the delivery of granular autograft or allograft into Spineology''s OptiMesh implant. Fill tube is identified with Spineology, catalog number and lot number printed directly onto the metal.

Recall number
Z-0677-05
Recalling firm
The Spineology Group
Product
Fill Tube,diverted, Catalog # 300-1002, is used for the delivery of granular autograft or allograft into Spineology''s OptiMesh implant. Fill tube is identified with Spineology, catalog number and lot number printed directly onto the metal.
Status
Terminated
Initiated
2005-01-28
Posted
2005-04-06
Terminated
2005-11-17
Reason
Three malfunctions, due to jamming of bone material in the fill tube occurred, which allowed the fill tube to advance more deeply than intended into the surgical site. Fill tubes are used in delivery of bone graft materials.
Root cause
Other
Action
A Product Update, January 2005 (Vol. 2) was e-mailed to all representatives and also sent by mail on January 28, 2005. Telephone calls to each distributor/representative were made by Spineology Sales & Marketing staff between January 28 and January 31. The Product Update informs customers that they will be receiving new tubes and new platens soon and that any tubes and platens currently in their possession should be returned to Spineology.
Quantity
3392
Distribution
11 consignees located in: Arkansas, Arizona, California, Georgia, Illinois, Michigan, Minnesota, North Carolina, New York, Oklahoma and United Kingdom
Lot, serial or model codes
Lot # UN02002, 71-1057, UN04003 and UN04004
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
3003179719
Firm city
Stillwater
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register