Skip to the content

FDA Medical Device Recalls (openFDA)

Cordis US Corp: MYNXGRIP 6F/7F, REF MX6721, Vascular Closure Device (VCD). Used to seal femoral arterial and femoral venous access sites

Recall number
Z-0675-2024
Recalling firm
Cordis US Corp
Product
MYNXGRIP 6F/7F, REF MX6721, Vascular Closure Device (VCD). Used to seal femoral arterial and femoral venous access sites
Status
Open, Classified
Initiated
2023-11-28
Posted
2024-01-05
Terminated
not on file
Reason
Cordis became aware that the label for the carton box containing 10 units is incorrect for some boxes of two lots of MynxGrip. The carton box from lot F2322902 (Item MX5021, MynxGrip 5F) as labeled as lot F2322903 (MX6721, MynxGrip 6/7F) and product from lot F2322903 (MX6721, MynxGrip 6/7F) was labeled as F2322902 (Item MX5021, MynxGrip 5F). The primary label on individual devices is correct.
Root cause
Labeling Change Control
Action
Cordis issued an Urgent MEDICAL DEVICE RECALL notice to its consignees on 11/28/2023 via via FEDEX. The notice explained the issue and requested the return of any affected product in inventory. For questions related to the recall and/or acknowledgement form that are not adequately addressed in the letter, please contact Cordis QA at: GMB-CordisFieldAction@cordis.com or call Cordis Customer Support 1-800-327-7714 Option 2, Monday through Friday from 8:00 AM to 8:00 PM EDT.
Quantity
70 units
Distribution
US: NC, WV, TN, MS, MI, OH, CA
Lot, serial or model codes
UDI/DI 10862028000403, Lot numbers: F2322903
510(k) numbers
not on file
PMA numbers
["P040044"]
Product code
MGB
Firm FEI
1016427
Firm city
Miami Lakes
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register